Dose response evaluation of a low density anoxic polymer gel dosimeter using MRI

نویسندگان

  • D. Shahbazi-Gahrouei Department of Medical Physics and Medical Engineering, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran
  • M. Gholami Department of Medical Physics and Medical Engineering, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran
  • T. Allahverdi Pourfallah Department of Biophysics and Biochemistry, School of Medicine, Mazandaran University of Medical Sciences, Sari, Iran
چکیده مقاله:

Background: The human body contains of different tissues and cavities with different physical and radiological properties. Most important among these are tissues and cavities that are radiologically different from water, including lungs, sinuses and bones. Gel dosimetry provides a unique feature to display dose distributions occurring in clinical radiation therapy in three dimensions. Materials and Methods: The low density polymer gel dosimeter is composed of 12% gelatin, 5% methacrylic acid, 0.15% sodium dodecyl sulfate, 10 mM THPC, and approximately 83% ultrapure deionized water. Post- preparation irradiation time for all samples was 5 hr. The time between irradiation and scanning for all gels experiments was 18 hr. The gel dosimeters were imaged using a 1.5 T clinical MRI scanner in a transmitter/receiver head coil. Results: There was a linear correlation between the doses and R2 responses from 0 to 12 Gy. However, above the 14 Gy probably due to saturation and or consumption of the monomers the dose response was reduced. The low-density gels had a mass density between 0.35 and 0.45 g.cm-3 and the CT values of about -650 to -750 Hounsfield units. These values are close to those of the normal human lung tissue, which ranges from -770 to -875 Hounsfield units. Conclusion: Increasing the gel temperature during rotation in the household mixer and probably reactions between the gelatin-free radicals and monomers led to a higher R2-background response.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

dose response evaluation of a low density anoxic polymer gel dosimeter using mri

background: the human body contains of different tissues and cavities with different physical and radiological properties. most important among these are tissues and cavities that are radiologically different from water, including lungs, sinuses and bones. gel dosimetry provides a unique feature to display dose distributions occurring in clinical radiation therapy in three dimensions. materials...

متن کامل

Basic investigation on performance of low-density polymer gel dosimeter

Background:  In this study a series of basic dosimetric properties of a low-density (LD) gel dosimeter are investigated. The dose response is studied regarding to linearity, sensitivity, dose-rate and energy dependence as well as lung tissue equivalence. Materials and Methods: The LD gel was made by mixing the polymer gel with expanded polystyrene spheres. Energy dependence was studied...

متن کامل

basic investigation on performance of low-density polymer gel dosimeter

background:  in this study a series of basic dosimetric properties of a low-density (ld) gel dosimeter are investigated. the dose response is studied regarding to linearity, sensitivity, dose-rate and energy dependence as well as lung tissue equivalence. materials and methods: the ld gel was made by mixing the polymer gel with expanded polystyrene spheres. energy dependence was studied at two d...

متن کامل

Dose integration and dose rate characteristics of a NiPAM polymer gel MRI dosimeter system

The normoxic polymer gel dosimeter based on N-isopropyl acrylamide (NiPAM) is a promising full 3D-dosimeter with high spatial resolution and near tissue equivalency. NiPAM gel samples were irradiated to different doses using a linear accelerator. The absorbed dose was evaluated using MRI and statistical significance of the analysed data was calculated. The analysis was carried out using an in-h...

متن کامل

Small field size dose-profile measurements using MAGIC- f polymer gel dosimeter and EBT3 film dosimeter

Introduction: In radiotherapy, treatment planning methods are evaluating and transforming every day, so precision in calculation and determination of dose delivery process is performing more important role in radiotherapy than past. In modern radiotherapy treatments, deramatically increase of small fields utilization, causes the polymeric gels has known as the most reliable dos...

متن کامل

Evaluation of MRI-based MAGIC polymer gel dosimeter in small photon fields

Background: Accurate small radiation field dosimetry is essential in modern radiotherapy techniques such as stereotactic radiosurgery (SRS) and intensity modulated radiotherapy (IMRT). Precise measurement of dosimetric parameters such as beam profile, percentage depth doses and output factor of these beams are complicated due to the electron disequilibrium and the steep dose gradients. In the p...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 13  شماره None

صفحات  243- 249

تاریخ انتشار 2015-07

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

کلمات کلیدی

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023