Determination of the optimum filter for qualitative and quantitative 99mTc myocardial SPECT imaging
نویسندگان
چکیده مقاله:
Background: Butterworth, Gaussian, Hamming, Hanning, and Parzen are commonly used SPECT filters during filtered back-projection (FBP) reconstruction, which greatly affect the quality and size accuracy of image. Materials and Methods: This study involved a cardiac phantom in which 1.10 cm thick cold defect was inserted into its myocardium wall and filled with 4.0 μCi/ml (0.148 MBq/ml) 99mTc concentration. The cardiac insert was then put into a cylindrical tank which was filled with 1.2 μCi/ml (0.044 MBq/ml) 99mTc concentration as background. 272 combinations of filter parameters were selected from those filters and applied on image. The measurements of count in myocardium, background, and defect regions of interest (ROI) were performed on each filtered image. Those measurements were then used to calculate contrast, signal-to-noise ratio (SNR), and defect size. For every filter, each criterion was graded (1 to 100) and then summed at their specific setting for total comparison. Results: The results show that, the different filter types produced myocardial image with different contrast, SNR, and defect size. For contrast and SNR, Gaussian filter was the best, while Parzen filter was the best in producing accurate defect size. However, Butterworth filter was found the best for trade off between contrast, SNR, and defect size accuracy. Conclusion: Selection of filter should consider the type of analysis, whether qualitatively or quantitatively. Qualitative analysis depends on image quality which is denoted by high contrast and SNR, thus Gaussian filter was suggested. Instead, the Butterworth filter was suggested for quantitative analysis as it is greatly dependent on both, image quality and size accuracy. Iran. J. Radiat. Res., 2009 6 (4): 173-182
منابع مشابه
determination of the optimum filter for qualitative and quantitative 99mtc myocardial spect imaging
background: butterworth, gaussian, hamming, hanning, and parzen are commonly used spect filters during filtered back-projection (fbp) reconstruction, which greatly affect the quality and size accuracy of image. materials and methods: this study involved a cardiac phantom in which 1.10 cm thick cold defect was inserted into its myocardium wall and filled with 4.0 μci/ml (0.148 mbq/ml) 99mtc conc...
متن کاملFilter selection for 99mTc-Sestamibi myocardial perfusion SPECT imaging [Persian]
Myocardial perfusion SPECT imaging with Tc99m-sestamibi is the most accurate non-invasive means of detecting coronary artery disease and assessing the severity of perfusion abnormalities in the patients with coronary stenosis. Though simple and straight forward the results produced by the technique is very much affected by the details being used. One of the main problems with the test, wh...
متن کاملDetermination of optimum filter in myocardial SPECT: A phantom study
Background: In myocardial perfusion SPECT imaging, images are degraded by photon attenuation, the distance-dependent collimator, detector response and photons scatter. Filters greatly affect quality of nuclear medicine images Materials and Methods: A phantom simulating heart left ventricle was built. About 1mCi of 99mTc was injected into the phantom. Images were taken from this phantom. Some fi...
متن کاملfilter selection for 99mtc-sestamibi myocardial perfusion spect imaging [persian]
myocardial perfusion spect imaging with tc99m-sestamibi is the most accurate non-invasive means of detecting coronary artery disease and assessing the severity of perfusion abnormalities in the patients with coronary stenosis. though simple and straight forward the results produced by the technique is very much affected by the details being used. one of the main problems with the test, which ha...
متن کاملRelationship between the optimum cut off frequency for Butterworth filter and lung-heart ratio in 99mTc myocardial SPECT
Background: We investigated whether the lungheart ratio parameter (LHR) can be used to identify the optimum cut off frequency for Butterworth filter in 99mTc myocardial SPECT imaging. Materials and Methods: This study involved a cardiac phantom system consisting of cardiac insert in which 1.10 cm cold defect was inserted into its myocardium wall and filled with 4.0 μCi/ml (0.148 MBq/ml...
متن کاملdetermination of optimum filter in myocardial spect: a phantom study
background: in myocardial perfusion spect imaging, images are degraded by photon attenuation, the distance-dependent collimator, detector response and photons scatter. filters greatly affect quality of nuclear medicine images materials and methods: a phantom simulating heart left ventricle was built. about 1mci of 99mtc was injected into the phantom. images were taken from this phantom. some fi...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ذخیره در منابع من قبلا به منابع من ذحیره شده{@ msg_add @}
عنوان ژورنال
دوره 6 شماره None
صفحات 173- 182
تاریخ انتشار 2009-03
با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023