Comparison of the Obtained Results from Two Calibration Methods in MRI Brachytherapy Normoxic Polymer Gel Dosimetry

نویسندگان

  • M. Allahverdi Assistant Professor, Dept. of Medical Physics, Faculty of Medicine, TUMS & Cancer Institute of Imam Hospital, Tehran, Iran.
  • M. Sedaghat Ph.D. Student in Medical Physics, University of Sherbrooke, Canada.
  • M.H Zahmatkesh Assistant Professor, Novin Medical Radiation Institute, Tehran, Iran.
  • R. Jaberi M.Sc. in Medical Physics, Cancer Institute, Imam Hospital, TUMS, Tehran, Iran.
  • Sh. Akhalghpoor Assistant Professor, Faculty of Medicine, TUMS & Novin Medical Radiation Institute, Tehran, Iran.
چکیده مقاله:

Introduction: Despite the fact that the clinical implementation of polymer gel dosimetry has been facilitated after the introduction of normoxic gels, there are still complications in its clinical routine use that are under investigation. In the current work, the feasibility of using a normoxic polymer gel dosimeter named MAGICA has been investigated for use in our clinical brachytherapy applications at ImamHospital and the deviations between the results obtained from two different calibration methods were investigated. Materials and Methods: The gel fabrication was performed at Novin Medical Radiation Institute and 4 Perspex walled phantoms designed especially for brachytherapy irradiations were filled with the same gel, together with 15 plastic calibration test tubes. The test tubes were irradiated with a range of known doses using a cobalt-60 teletherapy unit and the phantoms were irradiated with two brachytherapy remote-after-loaders using Cs137 and Co60 brachytherapy sources according to the treatment planning data. Imaging was performed with a Multi-spin-echo protocol using a Siemens 1.5T MRI machine. Image processing was performed in MATLAB environment to extract R2 maps of the dose distributions. Gel fabrication, irradiation and imaging were performed at three different centers, so the dosimeters and test tubes were transferred from one center to another under non-uniform conditions. R2 values of the first dosimeter which was irradiated with a single cesium source were extracted from different distances from the source along a horizontal profile and correlated to the dose values at the same distances in treatment planning data. This was used as the second calibration method to extract the absolute measured values of the other three dosimeters. Diagrams of the absorbed dose versus distance were plotted for each dosimeter and the results obtained from the two calibration methods were compared. Results: Results of the measured values that were calibrated with the test tubes show a 3 to 6 Gy dose difference in plateaus and more than 2 mm distance to agreement gap in steep dose gradients. However, the results of using the second calibration method show at most a 0.36 Gy dose difference in plateaus and less than 2 mm distance to agreement gap in steep dose gradients. Discussion and Conclusion: As the gel fabrication, irradiation and imaging conditions were the same for all the gel dosimeters and the calibration tubes, the unexpected deviations between the results of the two calibration methods deserves more investigations. It seems plausible to attribute this deviation to the difference in the size of calibration test tubes and original dosimeters. From a practical point of view, as the origin of this effect has not yet been investigated, it would be best to use a calibration phantom with the same size and characteristics as the original dosimeters.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

comparison of the obtained results from two calibration methods in mri brachytherapy normoxic polymer gel dosimetry

introduction: despite the fact that the clinical implementation of polymer gel dosimetry has been facilitated after the introduction of normoxic gels, there are still complications in its clinical routine use that are under investigation. in the current work, the feasibility of using a normoxic polymer gel dosimeter named magica has been investigated for use in our clinical brachytherapy applic...

متن کامل

Evaluation of Dose Distribution Accuracy in HDR Brachytherapy of Esophagus Cancer Based on MRI Normoxic Polymer Gel Dosimetry

Introduction: The purpose of this work was to study the ability of MRI normoxic polymer gel dosimetry for evaluating the dose distribution in HDR brachytherapy of esophagial cancer at Imam Reza brachytherapy center (Mashhad, Iran). Materials and Methods: Initially, 2liters of normoxic gel (MAGIC) was fabricated and then poured into 12 calibration test tubes and placed in a perspex walled phanto...

متن کامل

evaluation of dose distribution accuracy in hdr brachytherapy of esophagus cancer based on mri normoxic polymer gel dosimetry

introduction: the purpose of this work was to study the ability of mri normoxic polymer gel dosimetry for evaluating the dose distribution in hdr brachytherapy of esophagial cancer at imam reza brachytherapy center (mashhad, iran). materials and methods: initially, 2liters of normoxic gel (magic) was fabricated and then poured into 12 calibration test tubes and placed in a perspex walled phanto...

متن کامل

Brachytherapy polymer gel dosimetry with xCT

ABSTRACTBackground: Polymer gels are an emerging new class of dosimeters which are being applied to the challenges of modern radiotherapy modalities. Research on gel dosimetry involves several scientific domains, one of which is the imaging techniques with which dose data is extracted from the dosimeters. In the current work, we present our preliminary results of investigating capability of X-r...

متن کامل

Uncertainty Analysis in MRI-based Polymer Gel Dosimetry

Background: Polymer gel dosimeters combined with magnetic resonance imaging (MRI) can be used for dose verification of advanced radiation therapy techniques. However, the uncertainty of dose map measured by gel dosimeter should be known. The purpose of this study is to investigate the uncertainty related to calibration curve and MRI protocol for MAGIC (Methacrylic and Ascorbic acid in Gelatin I...

متن کامل

Verification of MRI MAGAT polymer gel dosimetry in clinical radiotherapy applications

Introduction: The objective of this work is to study the ability of MRI normoxic polymer gel dosimetry system to be used to verify the calculated dose distribution, in a simple geometry, in clinical radiotherapy. Materials and Methods:  Initially, 2 liters of the normoxic MAGAT polymer gel was prepared and then poured into two cylindrical Perspex phantom (simu...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 3  شماره 1

صفحات  11- 22

تاریخ انتشار 2006-03-01

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023