A Monte Carlo Simulation of Photon Beam Generated by a Linear Accelerator

نویسندگان

  • A. Hashemian Assistant Professor, Medical Physics Research Center, Bu-Ali Research Institute, Mashhad University of Medical Sciences, Mashhad, Iran.
  • H. Miri Assistant Professor, Nuclear Physics Dept., Ferdowsi University, Mashhad, Iran.
  • H. Saberi Ph.D. Student in Medical Physics, Mashhad university of Medical Sciences, Mashhad, Iran
  • M. H. Bahreyni Toosi Professor, Medical Physics Research Center, Bu-Ali Research Institute, Mashhad University of Medical Sciences, Mashhad, Iran.
  • M. Momen Nezhad Ph.D. Student in Medical Physics, Mashhad university of Medical Sciences, Mashhad, Iran
  • M. T. Bahreyni Toosi Professor, Medical Physics Research Center, Bu-Ali Research Institute, Mashhad University of Medical Sciences, Mashhad, Iran.
  • R. Salek Assistant Professor, Radiation Oncology Dept., Imam Reza Hospital, Mashhad University of Medical Sciences, Mashhad, Iran.
چکیده مقاله:

ntroduction:  Monte  Carlo  simulation  is  the  most  accurate  method  of  simulating  radiation  transport  and  predicting doses at different points of interest in radiotherapy. A great advantage of the Monte Carlo method  compared  to  the  deterministic  methods  is  the  ability  to  deal  accurately  with  any  complex  geometry.  Its  disadvantage is the extremely long computing time required to obtain a dose distribution with good statistical  accuracy.  Materials and Methods: The MCNP-4C Monte Carlo code was used to simulate a 9 MV photon beam from  a Neptun 10PC linear accelerator. The accelerator was modeled as a complete unit consisting of a target, exit  window, initial collimator, primary collimator, flattening filter, monitor chamber and secondary collimator.  The geometrical details and the composition of each component was either obtained from the manufacturer or  was  directly  measured.  The  simulation  of  the  source  was  performed  in  a  two  step  process.  Initially,  the  electron source was defined. Secondly, the bremsstrahlung energy spectra and the fluence distribution at the  scoring  planes  were  used  to  define  the  photon  source.  The  simulated  electron  beam  energy  followed  a  Gaussian distribution, with FWHM equal to 12% in nominal energy. The used intensity distribution of the  electron beam also followed a Gaussian distribution with a FWHM equal to 0.34 cm. To compute the photon  beam data a 50 × 50 × 40 cm 3  water phantom located at SSD = 100 cm was simulated. The depth dose and  the dose profile curves were calculated for four different field sizes (5×5, 10×10, 20×20 and 30×30 cm 2 ) and  compared against the measured values. The low-energy cut-off for the photons and electrons was 10 and 500  KeV, respectively. The measurements were carried out by using a Scanditronix dose scanning system and a  0.12 cm 3  RK ionization chamber.  Results: To verify the simulated model, the calculated Monte Carlo dose data were compared against the  corresponding measured values. The energy spectra and the angular distribution of the x-ray beam generated  by the Neptun 10PC linac was examined. The result showed an efficiency of about 73% for the production of  bermsstrahlung photon by the target. The agreement between the calculated and the measured depth dose and  the dose profile was generally better than 2% for all the fields.  Discussion and Conclusion: The simulation of the Neptun 10PC linac performed in this work is capable of  computing the depth dose data and the beam profiles in water phantom for all the predefined fields including  5×5, 10×10, 20×20 and 30×30 cm 2 . Therefore, it can be concluded that MCNP-4C is a suitable tool for the  dose calculation in radiotherapy. The simulated linac machine and the resulting data can be used to predict  the dose distribution in all complex fields. 

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

a monte carlo simulation of photon beam generated by a linear accelerator

ntroduction:  monte  carlo  simulation  is  the  most  accurate  method  of  simulating  radiation  transport  and  predicting doses at different points of interest in radiotherapy. a great advantage of the monte carlo method  compared  to  the  deterministic  methods  is  the  ability  to  deal  accurately  with  any  complex  geometry.  its  disadvantage is the extremely long computing time r...

متن کامل

Monte Carlo Simulation of a Linear Accelerator and Electron Beam Parameters Used in Radiotherapy

Introduction: In recent decades, several Monte Carlo codes have been introduced for research and medical applications. These methods provide both accurate and detailed calculation of particle transport from linear accelerators. The main drawback of Monte Carlo techniques is the extremely long computing time that is required in order to obtain a dose distribution with good statistical accuracy. ...

متن کامل

monte carlo simulation of a linear accelerator and electron beam parameters used in radiotherapy

introduction: in recent decades, several monte carlo codes have been introduced for research and medical applications. these methods provide both accurate and detailed calculation of particle transport from linear accelerators. the main drawback of monte carlo techniques is the extremely long computing time that is required in order to obtain a dose distribution with good statistical accuracy. ...

متن کامل

Benchmarking of Monte Carlo model of 6 Mv photon beam produced by Siemens Oncor® linear accelerator: determination of initial electron beam parameters in comparison with measurement

Introduction: The aim of this study was to investigate the initial electron beam parameters for Monte Carlo model of 6MV photon beam produced by Siemens Oncor® linear accelerator.   Materials and Methods: In this study, the EGSnrc Monte Carlo user codes BEAMnrc and DOSXYZnrc were used. The beamnrc code were used for modelling of a 6 MV photon beam produced by...

متن کامل

Localization of electron virtual SSD in a Siemens-Primus linear accelerator: Comparison of measurements with Monte Carlo simulation

Introduction: Because of importance of impact of the Source to Surface Distance (SSD) in determining of monitor unit for electron-therapy, it is essential to know the Virtual Source Position (VSP) for electron beam for a linear therapeutic accelerator for each energy and field size. , especially using the Khan method (photo distance squared method) And compare the results with...

متن کامل

Monte carlo simulation of varian clinac iX 10 MV photon beam for small field dosimetry

Background: The lack of lateral electronic disequilibrium (LED) becomes a main problem in small field. This factor affects the dose in target volume cannot predict correctly. In addition, utilization of high-energy linear accelerator (10 MV) can emit some unwanted particles (electron contamination). Therefore, the aim of this study was to characterize head linear accelerator (linac) Varian Clin...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 2  شماره 2

صفحات  3- 12

تاریخ انتشار 2005-06-01

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023