S. Senthilkumar
Department of Radiotherapy, Madurai Medical College and Govt. Rajaji Hospital, Maduari-625 020, India
[ 1 ] - Design of homogeneous and heterogeneous human equivalent thorax phantom for tissue inhomogeneity dose correction using TLD and TPS measurements
Background: The purpose of this study is to fabricate inexpensive in-house low cost homogeneous and heterogeneous human equivalent thorax phantom and assess the dose accuracy of the Treatment Planning Systems (TPS) calculated values for different lung treatment dosimetery. It is compared with Thermoluminescent Dosimeter (TLD) measurement. Materials and Methods: Homogeneous and heterogeneous tho...
[ 2 ] - Fabrication of low cost in-house slab homogeneous and heterogeneous phantoms for lung radiation treatment
Background: The heterogeneous composition of the human body presents numerous tissue types and cavities with widely differing radiologic properties. The aim of the present work was to develop a low cost homogeneous and heterogeneous phantom and the absorbed dose were measured by ionization chamber for different radiotherapy treatment techniques and compared with treatment planning syst...