Treatment couch positioning uncertainties using an EPID-based method

Authors

  • Ali Asghar Mowlavi Physics Department, School of Sciences, Hakim Sabzevari University, Sabzevar, Iran. ICTP, Associate federation Scheme, Medical Physics Field, Trieste, Italy.
  • Azam Afzalifar Physics Department, School of Sciences, Hakim Sabzevari University, Sabzevar, Iran.
  • Mohammad Mohammadi Department of Medical Physics, Royal Adelaide Hospital, Adelaide, SA 5000 AUSTRALIA. School of Physical Sciences, the University of Adelaide, SA 5005 AUSTRALIA. Department of Medical Physics, Reza Radiotherapy and Oncology Center, Mashhad, Iran
  • Sara Abdollahi Department of Medical Physics, Reza Radiotherapy and Oncology Center, Mashhad, Iran.
Abstract:

Introduction: The accuracy of patient positioning plays an important role for radiotherapy tasks. Short and automated Quality Assurance (QA) programs are becoming one of the challenging tasks in Radiotherapy. The current study, investigates the accuracy of treatment couch positioning with four degrees of freedom (4DoF) using a fast and accurate method based on an image acquired using an electronic portal imaging device (EPID). Materials and Methods: A Siemens medical linear accelerator equipped with an a-Si EPID was used to capture portal images. An accurate EPID QA phantom (PTW EPID QC Phantom®) and a proper in-house algorithm were used. The phantom has different sections to assess EPID image quality. Preliminary tests including Mechanical performance of the EPID positioning, EPID sensitivity and the code response from image processing point of view were investigated through several deviations deliberately created to evaluate translational and rotational displacements of treatment couch. Loading effect on treatment couch was also investigated. Results: The results of EPID preliminary tests and in-house code reliability were within tolerances expected. The mean deviations of tests for (4DoF) treatment couch were 0.25°±0.10˚, 0.428±0.112 mm, 0.428±0.112 mm and 0.286±0.112 mm for rotational, longitudinal, lateral and vertical shifts, respectively. The results of incline caused by a 130kg loading on the couch showed that the proposed method is a reliable and fast to find the uncertainties occurred for treatment couch positioning. Conclusion: As a part of linac QA program, treatment couch tests including translational and rotational motions under unloaded/loaded conditions were performed. The corresponding errors for table positioning including minimum detectable rotational and translational displacements, and couch sagging effect under loaded conditions were evaluated using an EPID. Creating one major test for the whole assessments for couch checks, fast and accurate and no labour intensive procedure, and electronic portal image instead of film and other imagers can be addressed as advantages of the method proposed.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

An Empirical Transmitted EPID Dosimetry Method using a Back-Projection Algorithm

Background: The present study aimed to introduce a rapid transmission dosimetry through an electronic portal-imaging device (EPID) to achieve two-dimensional (2D) dose distribution for homogenous environments. Material and Methods: In this Phantom study, first, the EPID calibration curve and correction coefficients for field size were obtained from EPID and ionization chamber. Second, the...

full text

Evaluation of a fast method of EPID-based dosimetry for IMRT and Comparison with 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT

Introduction: Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) and VMAT QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in ...

full text

WE-G-BRB-02: MU-EPID an EPID Based Tool for IMRT Quality Assurance.

PURPOSE A software program (MU-EPID) has been developed to perform patient specific IMRT pre-treatment QA verification using an electronic portal imaging device. METHODS The software converts measured images of intensity modulated beams delivered to an EPID, into fluence maps that can be imported in the treatment planning system. The dose can then be calculated in the patient anatomy and comp...

full text

Development and implementation of an EPID‐based method for localizing isocenter

The aim of this study was to develop a phantom and analysis software that could be used to quickly and accurately determine the location of radiation isocenter to an accuracy of less than 1 mm using the EPID (Electronic Portal Imaging Device). The proposed solution uses a collimator setting of 10 × 10 cm2 to acquire EPID images of a new phantom constructed from LEGO blocks. Images from a number...

full text

An Indoor Positioning Method for Smartphones Using Landmarks and PDR †

Recently location based services (LBS) have become increasingly popular in indoor environments. Among these indoor positioning techniques providing LBS, a fusion approach combining WiFi-based and pedestrian dead reckoning (PDR) techniques is drawing more and more attention of researchers. Although this fusion method performs well in some cases, it still has some limitations, such as heavy compu...

full text

Proposing a Framework for Strategic Positioning Using an Integrated Method

Thisarticleis to determine thecompany'sstrategyanditsstrategicpositioning. At first, the most influential internal and external elements were detected with the help of the techniques of strategy formulation. Then using of Strengths, Weaknesses, Opportunities and Threats (SWOT) matrix, we formulated the primary organizational strategies. Using of fuzzy TOPSIS, priorityscore of strategies isspeci...

full text

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 15  issue Special Issue-12th. Iranian Congress of Medical Physics

pages  226- 226

publication date 2018-12-01

By following a journal you will be notified via email when a new issue of this journal is published.

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023