The use of EBT3 film and Delta4 for the dosimetric verification of EclipseTM treatment planning system in a heterogeneous chest phantom: an IMRT technique
Authors
Abstract:
Background: This study aimed to evaluate the dose calculation accuracy of EclipseTM treatment planning system (TPS) in a heterogeneous chest phantom with the intensity modulated radiotherapy (IMRT) technique using EBT3 film and Delta4. Materials and Methods: Two IMRT plans (A and B) were prepared for radiotherapy of the heterogeneous chest phantom. Plan A was between the left lung and the surrounding soft tissue and plan B was carried out on the soft tissue. EBT3 film and Delta4 were used for dose measurement in the phantom. EclipseTM TPS was also used for dose calculation. Finally, the gamma index values of the TPS with film and TPS with Delta4 were obtained. A 95% passing rate of gamma index with the passing criterion of 3mm/3% and a dose threshold of 20% as the standard criterion was considered in this study. Furthermore, the passing rates of gamma indices of the film and Delta4 were compared with each other via Bland-Altman analysis. Results: The mean passing rate of gamma index with standard passing criterion between the TPS calculations and film measurements was 96.95±0.22%, while it was equal to 97.7±0.56% and 98.45±0.21% between the TPS calculations and 2D and 3D Delta4 measurements, respectively. Additionally, the differences between the passing rates of gamma indices of the film and Delta4 were less than 5%. Conclusion: The findings demonstrate that the accuracy of dose calculations of EclipseTM TPS in a heterogeneous chest phantom with the IMRT technique is within the standard passing criterion. Furthermore, it can be concluded that there is a good agreement between the film and Delta4, as IMRT QA devices.
similar resources
مقایسه عملکرد فیلم گف کرومیک EBT3و فانتوم Delta4® در تکنیک IMRT محیط ناهمگن قفسه سینه
Introdution: In advanced radiotherapy techniques such as intensity modulated radiation therapy (IMRT), the quality assurance (QA) is essential. This study aimed to compare the performance between GafchromicTM EBT3 film and Delta4® phantom (2D and 3D) in heterogeneous chest phantom using IMRT technique. Methods: In this experimental study, two IMRT plans (A and B) were prepared for radiothe...
full textDose calculations accuracy of TiGRT treatment planning system for small IMRT beamlets in heterogeneous lung phantom
Background: Accurate dose calculations in small beamlets and lung material have been a great challenge for most of treatment planning systems (TPS). In the current study, the dose calculation accuracy of TiGRT TPS was evaluated for small beamlets in water and lung phantom by comparison to Monte Carlo (MC) calculations. Materials and Methods: The head of Siemens Oncor-impression linac...
full textEvaluation of the Delta4 phantom for IMRT and VMAT verification.
The Delta(4) diode array phantom (Scandidos, Uppsala, Sweden) was evaluated for verification of segmental intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) on an Elekta linear accelerator (Crawley UK). The device was tested for angular sensitivity by irradiating it from 36 different gantry angles, and the responses of the device to various step-and-shoot s...
full textDosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification
Radiochromic film has become an important tool to verify dose distributions in highly conformal radiation therapy such as IMRT. Recently, a new generation of these films, EBT3, has become available. EBT3 has the same composition and thickness of the sensitive layer of the previous EBT2 films, but its symmetric layer configuration allows the user to eliminate side orientation dependence, which i...
full textan investigation of the types of text reduction in subtitling: a case study of the persian film gilaneh with english subtitles
چکیده ندارد.
15 صفحه اولthe use of appropriate madm model for ranking the vendors of mci equipments using fuzzy approach
abstract nowadays, the science of decision making has been paid to more attention due to the complexity of the problems of suppliers selection. as known, one of the efficient tools in economic and human resources development is the extension of communication networks in developing countries. so, the proper selection of suppliers of tc equipments is of concern very much. in this study, a ...
15 صفحه اولMy Resources
Journal title
volume 17 issue 2
pages 355- 361
publication date 2019-04
By following a journal you will be notified via email when a new issue of this journal is published.
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023