Penumbra Measurements and Comparison of In-House and Standard Circular Cones by the Gafchoromic Film, Pinpoint Ion Chamber, and MCNPX Monte Carlo Simulation

Authors

  • Ehsan Mohammadi Radiotherapy Oncology Research Center, Cancer Institute, Tehran University of Medical Sciences, Tehran, Iran
  • Ghazale Geraily Department of Medical Physics, Tehran University of Medical Science, Tehran, Iran
  • hassan nedaie radiotherapy oncology department, cancer research centre, Tehran university of medical sciences,Tehran,Iran
  • Mahbod Esfehani Radiotherapeutic Oncology Department of Cancer Institute, Tehran, Iran
  • mansur naderi Radiotherapy Oncology Department, Cancer Research Centre, Tehran University of Medical Sciences, Tehran, Iran
  • mohsen hassani Department of Radiotherapy Physics, Cancer Research Centre, cancer Institute, Tehran University of Medical Sciences, Tehran, Iran
  • Sareh Tajiki Radiotherapy Oncology Research Center, Cancer Institute, Tehran University of Medical Sciences,Tehran, Iran.
Abstract:

Introduction: Penumbra is an important property of the radiation beam to obtain a suitable margin surrounding the target volume. Therefore, the precise penumbra width determination in stereotactic radiotherapy is necessary for treatment planning. This study aimed to compare the obtained results of penumbra width by in-house and standard circular cones by different dosimeters, as well as evaluating the function of EBT3 for dosimetric properties of the small field radiation. Material and Methods: Different circular cones were mounted on the head of the accelerator to produce 12, 20, and 40 mm field sizes at isocenter. Dosimetric measurements were performed with the EBT3 film, PinPoint ion chamber. Afterwards, MCNPX Monte Carlo simulation was used to evaluate the dosimetric parameters. Results: According to the obtained results, the penumbra width was increased by larger diameters of circular cones. The obtained measured data by PinPoint ion chamber showed a larger penumbra width compared to those calculated by Monte Carlo at all field sizes. The gamma index analysis revealed  distance-to-agreement and dose-difference of 2 mm /2%/ at all points. The results of this study showed that source to diaphragm distance had a major role in penumbra size determination of small field dosimetry with PinPoint ion chamber, EBT3 film, and Monte Carlo simulation. Conclusion: As findings of this study reported, EBT3 films are reliable detectors for relative dosimetry due to high spatial resolution for small field sizes. Furthermore, they can be used for measuring beam profile and percentage depth dose curves.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

Small photon field dosimetry using EBT2 Gafchromic film and Monte Carlo simulation

Background: Small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. Accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. Hence the EBT2 (external beam therapy) Gafchromic film was used for dosimetry of sm...

full text

Comparison of measured and Monte Carlo calculated dose distributions from “circular collimators” for radiosurgical beams

Background: Stereotactic radiosurgery is an important clinical tool for the treatment of small lesions in the brain, including benign conditions, malignant and localized metastatic tumors. A dosimetry study was performed for Elekta ‘Synergy S ’ as a dedicated Stereotactic radiosurgery unit, capable of generating circular radiation fields with diameters of 1-5 cm at isocentre using the BEAM/EGS4...

full text

Design and Simulation of Photoneutron Source by MCNPX Monte Carlo Code for Boron Neutron Capture Therapy

Introduction Electron linear accelerator (LINAC) can be used for neutron production in Boron Neutron Capture Therapy (BNCT). BNCT is an external radiotherapeutic method for the treatment of some cancers. In this study, Varian 2300 C/D LINAC was simulated as an electron accelerator-based photoneutron source to provide a suitable neutron flux for BNCT. Materials and Methods Photoneutron sources w...

full text

assessment of deep word knowledge in elementary and advanced iranian efl learners: a comparison of selective and productive wat tasks

testing plays a vital role in any language teaching program. it allows teachers and stakeholders, including program administrators, parents, admissions officers and prospective employers to be assured that the learners are progressing according to an accepted standard (douglas, 2010). the problems currently facing language testers have both practical and theoretical implications but the first i...

the survey of the virtual higher education in iran and the ways of its development and improvement

این پژوهش با هدف "بررسی وضعیت موجود آموزش عالی مجازی در ایران و راههای توسعه و ارتقای آن " و با روش توصیفی-تحلیلی و پیمایشی صورت پذیرفته است. بررسی اسنادو مدارک موجود در زمینه آموزش مجازی نشان داد تعداد دانشجویان و مقاطع تحصیلی و رشته محل های دوره های الکترونیکی چندان مطلوب نبوده و از نظر کیفی نیز وضعیت شاخص خدمات آموزشی اساتید و وضعیت شبکه اینترنت در محیط آموزش مجازی نامطلوب است.

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 16  issue 3

pages  232- 240

publication date 2019-05-01

By following a journal you will be notified via email when a new issue of this journal is published.

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023