Monte Carlo Study of the Effect of Backscatter Materail Thickness on 99mTc Source Response in Single Photon Emission Computed Tomography
Authors
Abstract:
Introduction SPECT projections are contaminated by scatter radiation, resulting in reduced image contrast and quantitative errors. Backscatter constitutes a major part of the scatter contamination in lower energy windows. The current study is an evaluation of the effect of backscatter material on FWHM and image quality investigated by Monte Carlo simulation. Materials and Methods SIMIND program was used for simulation of a Siemen’s dual-head variable angle scintillation gamma camera. Planar and SPECT scanning of a 99mTc source and a Jaszczak phantom for varying thicknesses of Perspex slabs, as a backscatter media, were analyzed using the photopeak and scatter windows. Simulated planar images and reconstructed tomographic images were evaluated qualitatively, by two nuclear medicine specialists, and quantitatively, by Structural Similarity (SSIM) Index. Results In the 99mTc photopeak window, no significant change in total counts due to backscatter material was measured. In the scatter windows, scattering was overestimated compared with a simulated backscatter free SPECT system. For instance, at a thickness of 10 cm, total counts of a 99mTc source detected in the 72 keV windows eventually doubled with increasing backscatter material, compared with the situation without backscatter material. The backscatter contribution plateaued when more than 7 cm of scatter material was placed but there were optimized results for a backscatter thickness of 4.5 cm. Better image quality for the thickness was confirmed by the results of eye interpretation and also by SSIM algorithm. Conclusion Backscatter should be taken into account, particularly in model-based scatter correction methods in SPECT for an accurate simulation system optimization.
similar resources
monte carlo study of the effect of backscatter materail thickness on 99mtc source response in single photon emission computed tomography
introduction spect projections are contaminated by scatter radiation, resulting in reduced image contrast and quantitative errors. backscatter constitutes a major part of the scatter contamination in lower energy windows. the current study is an evaluation of the effect of backscatter material on fwhm and image quality investigated by monte carlo simulation. materials and methods simind program...
full textImage Optimization in Single Photon Emission Computed Tomography by Hardware Modifications with Monte Carlo Simulation
Introduction: In Single Photon Emission Computed Tomography (SPECT), the projection data used for image reconstruction are distorted by several factors, including attenuation and scattering of gamma rays, collimator structure, data acquisition method, organ motion, and washout of radiopharmaceuticals. All these make reconstruction of a quantitative SPECT image very difficult. Simulation of a SP...
full textimage optimization in single photon emission computed tomography by hardware modifications with monte carlo simulation
introduction: in single photon emission computed tomography (spect), the projection data used for image reconstruction are distorted by several factors, including attenuation and scattering of gamma rays, collimator structure, data acquisition method, organ motion, and washout of radiopharmaceuticals. all these make reconstruction of a quantitative spect image very difficult. simulation of a sp...
full textDetermination of the Energy Windows for the Triple Energy Window Scatter Correction Method in Gadolinium-159 Single Photon Emission Computed Tomography Using Monte Carlo Simulation
Introduction: In radionuclide imaging, object scatter is one of the major factors leading to image quality degradation. Therefore, the correction of scattered photons might have a great impact on improving the image quality. Regarding this, the present study aimed to determine the main and sub-energy windows for triple energy window (TEW) scatter correction method usin...
full textA study on differentiation of extra-cardiac activity by Slit Slat collimation in Single Photon Emission Computed Tomography
Introduction: Myocardial perfusion SPECTby 99mTc-Sestamibi and 99mTc-Tetrofosmin radiopharmaceuticals usually presents a false significant increase in the radiotracer uptake in the inferior myocardium due to the uptake in organs such as liver, bowel, stomach and biliary system. The present study evaluated a suitable Slit angle for differentiating extra-car...
full textthe effect of consciousness raising (c-r) on the reduction of translational errors: a case study
در دوره های آموزش ترجمه استادان بیشتر سعی دارند دانشجویان را با انواع متون آشنا سازند، درحالی که کمتر به خطاهای مکرر آنان در متن ترجمه شده می پردازند. اهمیت تحقیق حاضر مبنی بر ارتکاب مکرر خطاهای ترجمانی حتی بعد از گذراندن دوره های تخصصی ترجمه از سوی دانشجویان است. هدف از آن تاکید بر خطاهای رایج میان دانشجویان مترجمی و کاهش این خطاها با افزایش آگاهی و هوشیاری دانشجویان از بروز آنها است.از آنجا ک...
15 صفحه اولMy Resources
Journal title
volume 10 issue 1
pages 69- 77
publication date 2013-03-01
By following a journal you will be notified via email when a new issue of this journal is published.
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023