Monte Carlo Simulation of Siemens Primus plus Linac for 6 and 18 MV Photon Beams
Authors
Abstract:
Objective: The aim of the present study is to simulate 6 MV and 18 MV photon beam energies of a Siemens Primus Plus medical linear accelerator (Linac) and to verify the simulation by comparing the results with the measured data.Methods: The main components of the head of Siemens Primus Plus linac were simulated using MCNPX Monte Carlo (MC) code. To verify the results, experimental data of percentage depth dose (PDD) and beam dose profile for 5 × 5 cm2, 10 × 10 cm2 and 20 × 20 cm2 field sizes were measured and compared with simulation results. Moreover, gamma function was used to compare the measurement and simulation data.Results: The results show a good agreement, within 1%, was observed between the data calculated by the simulations and those obtained by measurement for 6 MV photon beam, while it was within 2% for 18 MV photon beam, except in the build-up region for both beams. Gamma index values were less than unity in most data points for all the mentioned energies and fields. To calculate the dose in the phantom, cells were selected in different modes, one of the modes due to the lack of dose gradient and overlapping, produced better results than others produce.Conclusion: There was good settlement between measured and MC simulation values in this research. The simulation programs can be used for photon modes of Siemens Primus Plus linac in conditions in which it is not possible to perform experimental measurements.
similar resources
Monte Carlo Simulation of Siemens Primus plus Linac for 6 and 18 MV Photon Beams
Objective The aim of the present study is to simulate 6 MV and 18 MV photon beam energies of a Siemens Primus Plus medical linear accelerator (Linac) and to verify the simulation by comparing the results with the measured data. Methods The main components of the head of Siemens Primus Plus linac were simulated using MCNPX Monte Carlo (MC) code. To verify the results, experimental data of perc...
full textdetermination of primary electron beam parameters in a siemens primus linac using monte carlo simulation
the monte carlo method can be used to describe any technique that approximates solutions to quantitative problems through statistical sampling. this method is considered to be the most accurate method for dose calculation in radiotherapy. for complete modeling of a linear accelerator, it is required that the manufactured information covers all data, but some data such as primary electron energy...
full textEstimation of Dosimetric Parameters based on KNR and KNCSF Correction Factors for Small Field Radiation Therapy at 6 and 18 MV Linac Energies using Monte Carlo Simulation Methods
Background: Estimating dosimetric parameters for small fields under non-reference conditions leads to significant errors if done based on conventional protocols used for large fields in reference conditions. Hence, further correction factors have been introduced to take into account the influence of spectral quality changes when various detectors are used in non-reference conditions at differen...
full textA Monte Carlo study on electron and neutron contamination caused by the presence of hip prosthesis in photon mode of a 15 MV Siemens PRIMUS linac
Several investigators have pointed out that electron and neutron contamination from high-energy photon beams are clinically important. The aim of this study is to assess electron and neutron contamination production by various prostheses in a high-energy photon beam of a medical linac. A 15 MV Siemens PRIMUS linac was simulated by MCNPX Monte Carlo (MC) code and the results of percentage depth ...
full textMonte carlo simulation of varian clinac iX 10 MV photon beam for small field dosimetry
Background: The lack of lateral electronic disequilibrium (LED) becomes a main problem in small field. This factor affects the dose in target volume cannot predict correctly. In addition, utilization of high-energy linear accelerator (10 MV) can emit some unwanted particles (electron contamination). Therefore, the aim of this study was to characterize head linear accelerator (linac) Varian Clin...
full textThe comparison between 6 MV Primus LINAC simulation output using EGSnrc and commissioning data
Introduction: Monte Carlo calculation method is considered to be the most accurate method for dose calculation in radiotherapy. The purpose of this research is comparison between 6 MV Primus LINAC simulation output with commissioning data using EGSnrc and build a Monte Carlo geometry of 6 MV Primus LINAC as realistically as possible. The BEAMnrc and DOSXYZnrc (EGSnrc package) M...