Monte Carlo investigation on precise dosimetry of HDR breast brachytherapy with Accuboost
Authors
Abstract:
Introduction: Accuboost is a HDR brachytherapy system in early stages breast cancer treatment. This device provides a completely noninvasive procedure with parallel-opposed radiation from two immobilizing peripheral applicators that caused it a preferred option of modalities to choose. In most commercial treatment planning systems, tissues are considered as a simple water phantom. Inaccuracies arised from geometry and compositions can lead to deviations in interested dose distribution. This study intends to simulate HDR breast brachytherapy with Accuboost device in order to consider requirement corrections to have an accurate treatment dosimetry. Materials and Methods: MCNP5 simulation is employed to model phantoms in dosimetric measurements. We tried to take geometrical corrections into account in three phantoms. Dose distributions as dose volume histogram parameters and isodose plots obtained from these phantoms are compared to those of water. Corrections are included using real breast tissue with ICRU-44 breast composition, chest defined with an air cavity as lung and a curved surface to introduce a more realistic contour for breast skin. Accuboost is simulated with radiations from eighteen HDR 192-Ir sources on each applicator with appropriate tungsten shields. Results: Differences in most regions with less than 1% to water and maximum value about 5 % have been observed in phantom with ICRU-44 breast material compared to the water. Lung presence have increased deviations a few percent but it has not significantly differed. In phantom with all corrections maximum differences are almost the same as both earlier and deviations with about 5-7% are found near the chest and around the breast skin not in contact to applicators. Conclusion: Based on the results, although discrepancies between corrected phantom and water phantom are in areas near the skin the main differences are in clinical accepted uncertainties and other points of tissue have shown accommodatian. Therefore essential correction may be not needed.
similar resources
Accuracy Evaluation of Oncentraâ„¢ TPS in HDR Brachytherapy of Nasopharynx Cancer Using EGSnrc Monte Carlo Code
Background: HDR brachytherapy is one of the commonest methods of nasopharyngeal cancer treatment. In this method, depending on how advanced one tumor is, 2 to 6 Gy dose as intracavitary brachytherapy is prescribed. Due to high dose rate and tumor location, accuracy evaluation of treatment planning system (TPS) is particularly important. Common methods used in TPS dosimetry are based on computat...
full textDosimetric characterization of round HDR 192Ir accuboost applicators for breast brachytherapy.
PURPOSE The AccuBoost brachytherapy system applies HDR 192Ir beams peripherally to the breast using collimating applicators. The purpose of this study was to benchmark Monte Carlo simulations of the HDR 192Ir source, to dosimetrically characterize the round applicators using established Monte Carlo simulation and radiation measurement techniques and to gather data for clinical use. METHODS Do...
full textThermoluminescent and Monte Carlo dosimetry of IR06‐P103d brachytherapy source
This work presents experimental dosimetry results for a new 103Pd brachytherapy seed, in accordance with the AAPM TG-43U1 recommendation that all new low-energy interstitial brachytherapy seeds should undergo one Monte Carlo (MC) and at least one experimental dosimetry characterization. Measurements were performed using TLD-GR200A circular chip dosimeters using standard methods employing thermo...
full textMonte Carlo dosimetry of a new 90Y brachytherapy source
PURPOSE In this study, we attempted to obtain full dosimetric data for a new (90)Y brachytherapy source developed by the College of Chemistry (Sichuan University) for use in high-dose-rate after-loading systems. MATERIAL AND METHODS The dosimetric data for this new source were used as required by the dose calculation formalisms proposed by the AAPM Task Group 60 and Task Group 149. The active...
full textMonte Carlo investigation of energy response of various detector materials in I125 and Yb169 brachytherapy dosimetry
Relative absorbed-dose energy response correction R for different detector materials in water, PMMA and polystyrene phantoms are calculated using Monte Carlo-based EGSnrc code system for ¹²⁵I and ¹⁶⁹Yb brachytherapy sources. The values of R obtained for ¹²⁵I source are 1.41, 0.92, 3.97, 0.47, 8.32 and 1.10, respectively, for detector materials LiF, Li₂B₄O₇ , A1₂O₃, diamond, silicon diode and ai...
full textComparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations
Background: Iodine brachytherapy sources with low photon energies have been widely used in treating cancerous tumors. Dosimetric parameters of brachytherapy sources should be determined according to AAPM TG-43U1 recommendations before clinical use. Monte Carlo codes are reliable tools in calculation of these parameters for brachytherapy sources. Materials and Methods: Dosimetric param...
full textMy Resources
Journal title
volume 15 issue Special Issue-12th. Iranian Congress of Medical Physics
pages 155- 155
publication date 2018-12-01