Monte Carlo Evaluation of Gamma Knife Dose Profile in Real Brain Phantom
Authors
Abstract:
Introduction The Gamma Knife system is designed solely for non-invasive treatment of brain disorders, and it benefits from stereotactic surgical techniques. Dose calculations required in the system are performed by GammaPlan code; in this code, brain tissue is considered uniform. In the present study, we evaluated the effect of Gamma Knife system on the obtained dose through simulating a real human brain phantom. Materials and Methods In this study, a Monte Carlo simulation code (MCNPX2.7) was employed to simulate Gamma Knife system. Brain tissue equivalent Snyder phantom and combinations were considered according to International Commission on Radiological Units (ICRU)-44 report. Results To ensure accuracy of the simulations, patient’s head was modeled by a spherical water phantom. At this point, the dosimetry parameters were compared with those obtained by the Monte Carlo code EGS4 and good consistency was observed (less than 7% difference). At the next stage, the above dosimetry parameters were compared with those obtained experimentally by polystyrene phantom and EDR2 dosimetry film and improved consistency was detected (less than 0.5% difference). Finally, the Snyder phantom, as the human brain, was simulated. The Full Width at Half Maximum (FWHM) and penumbra decreased by 4.7% and 18%, respectively. Moreover, an isocenter dose reduction of 30-40%, compared to the water phantom, was noted. Conclusion The calculation of the real phantom showed that water and polystyrene could function similarly, while evaluating dosimetry parameters in the Gamma Knife system; thus, water and polystyrene are not appropriate phantom matters for this purpose.
similar resources
monte carlo evaluation of gamma knife dose profile in real brain phantom
introduction the gamma knife system is designed solely for non-invasive treatment of brain disorders, and it benefits from stereotactic surgical techniques. dose calculations required in the system are performed by gammaplan code; in this code, brain tissue is considered uniform. in the present study, we evaluated the effect of gamma knife system on the obtained dose through simulating a real h...
full textEvaluation of Organ Specific peripheral dose for Gamma knife 4C based on Monte Carlo
Introduction: Stereotactic Gamma Knife radiosurgery has been widely used for treating brain tumors. The scattered radiation outside of treatment field (peripheral dose) can induce the secondary cancer to specific organ. This paper investigated the absorbed dose to eyes, thyroid, heart, lung, breast and colon using a Monte Carlo technique for Mird phantom. We also study the ef...
full textEvaluation of the Effects of Inhomogeneities on Dose Profiles Using Polymer Gel Dosimeter and Monte Carlo Simulation in Gamma Knife
Introduction Polymer gel dosimeters offer a practical solution to 3D dose verification for conventional radiotherapy as well as intensity-modulated and stereotactic radiotherapy. In this study, EGSnrc calculated and PAGAT polymer gel dosimeter measured dose profiles from single shot irradiation with 18 mm collimator of Gamma Knife in homogeneous and inhomogeneous phantoms were compared with eac...
full textevaluation of the effects of inhomogeneities on dose profiles using polymer gel dosimeter and monte carlo simulation in gamma knife
introduction polymer gel dosimeters offer a practical solution to 3d dose verification for conventional radiotherapy as well as intensity-modulated and stereotactic radiotherapy. in this study, egsnrc calculated and pagat polymer gel dosimeter measured dose profiles from single shot irradiation with 18 mm collimator of gamma knife in homogeneous and inhomogeneous phantoms were compared with eac...
full textChoice of phantom materials for dosimetry of Leksell Gamma Knife unit: a Monte Carlo study.
In calculations for the Leksell Gamma Knife, GammaPlan employs a tissue equivalent material without the presence of a skull bone, while dosimetry work is based on a polystyrene phantom. The compatibility of these dose distributions is uncertain. The Monte Carlo technique was employed to determine the radial dose distributions from a single 14 mm collimator helmet in 160 mm diam phantoms with di...
full textMonte Carlo calculation of single-beam dose profiles used in a gamma knife treatment planning system.
The accuracy of single-beam dose profiles used in the algorithm of the Gamma Knife treatment planning system (Leksell GammaPlan) is verified. EGS4 Monte Carlo calculation was employed to calculate the dose distributions of single-beams in a spherical water phantom with diameter 160 mm. The beams were directed to the center of the phantom. Collimators of 4, 8, 14, and 18 mm sizes were studied. T...
full textMy Resources
Journal title
volume 13 issue 1
pages 1- 7
publication date 2016-03-01
By following a journal you will be notified via email when a new issue of this journal is published.
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023