Investigation of the Entrance Surface Dose and Dose to Different Organs in Lumbar Spine Imaging

Authors

  • R Faghihi
  • S Sina Radiation Research Center, Shiraz University, Shiraz, Iran
Abstract:

Background: Dose assessment using proper dosimeters is especially important in radiation protection optimization and imaging justification in diagnostic radiology.Objective: The aim of this study is to obtain the Entrance Skin Dose (ESD) of patients undergoing lumbar spine imaging using two thermoluminescence dosimeters TLD-100 (LiF: Mg, Ti) and GR-200 (LiF: Mg, Cu, P) and also to obtain the absorbed dose to different organs in lumbar spine imaging with several views.Methods: To measure the ESD values of the patients undergoing lumbar spine imaging, the two TLD types were put on their skin surface. The ESD values for different views of lumbar spine imaging were also measured by putting the TLDs at the surface of the Rando phantom. Several TLD chips were inserted inside different organs of Rando phantom to measure the absorbed dose to different organs in lumbar spine imaging.Results: The results indicate that there is a close agreement between the results of the two dosimeters. Based on the results of this experiment, the ESD dose of the 16 patients included in this study varied between 2.71 mGy and 26.29 mGy with the average of 11.89 mGy for TLD-100, and between 2.55 mGy and 27.41 mGy with the average of 12.32 mGy for GR-200 measurements. The ESDs obtained by putting the two types of TLDs at the surface of Rando phantom are in close agreement.Conclusion: According to the results, the GR200 has greater sensitivity than the TLD-100.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

investigation of the entrance surface dose and dose to different organs in lumbar spine imaging

background: dose assessment using proper dosimeters is especially important in radiation protection optimization and imaging justification in diagnostic radiology. objective: the aim of this study is to obtain the entrance skin dose (esd) of patients undergoing lumbar spine imaging using two thermoluminescence dosimeters tld-100 (lif: mg, ti) and gr-200 (lif: mg, cu, p) and also to obtain the a...

full text

Effects of Kilovoltage on Image Quality and Entrance Surface Dose in Lumbar Spine Digital Radiography

Introduction: Digital radiography possesses a wide dynamic range and has a major advantage in producing an acceptable image of diagnostic value even though overexposure occurs. Lumbar spine (LS) radiography is the most common examinations that gives high radiation dose to patients and accounts for the highest collective population dose of any conventional radiographic ...

full text

Evaluation of the Entrance Surface Dose (ESD) and Radiation Dose to the Radiosensitive Organs in Pediatric Pelvic Radiography

Background Patients' dosimetry is crucial in order to enhance radiation protection optimization and to deliver low radiation dose to the patients in a radiological procedure. The aim of this study was to assess the entrance surface dose (ESD) and radiation dose to the radiosensitive organs in pediatric pelvic radiography. Materials and Methods The studied population included 98 pediatric patien...

full text

Average Glandular Dose and Entrance Surface Dose in Mammography

In general, dose evaluation for mammography uses average glandular dose (AGD). In mammography, entrance surface dose (ESD) is high, which is well known, but actual clinical data of ESD are not reported so much. This study investigated AGD and ESD for patients whose doses were estimated from 200 mammography examinations using European Organization for Quality-Assured Breast Screening and Diagnos...

full text

Image Quality and Entrance Surface Dose Evaluation of Lateral Cervical Spine: A Study Using Grid and Non-Grid Techniques

Introduction: The purpose of this study is to investigate the effects of grid and non-grid techniques in the lateral cervical spine radiography on image quality and entrance surface dose (ESD). Although image quality and radiation doses have been studied by researchers, there is still a dearth of information on image quality and patient dose with different techniques. Material and Methods: The...

full text

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 4  issue 4

pages  -

publication date 2014-12-01

By following a journal you will be notified via email when a new issue of this journal is published.

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023