Investigation of LiF, Mg and Ti (TLD-100) Reproducibility
Authors
Abstract:
LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the values of Element Correction Coefficient (ECC) were obtained 4 times by irradiating them to doses of 81 mGy (two times), 162mGy and 40.5mGy. Results of this study indicate that the average reproducibility of ECC calculation for 40 TLDs is 1.5%, while these values for all chips do not exceed 5%.
similar resources
investigation of lif, mg and ti (tld-100) reproducibility
lif, mg and ti cubical tld chips (known as tld-100) are widely used for dosimetry purposes. the repeatability of tl dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mgy) with 662kev photons of cs-137. a group of 40 cubical tld chips was randomly selected from a batch and the values of element correction coefficient (ecc) were obtained 4 times by irradiating them to doses...
full textتولید دزیمتر LiF: Mg, Ti و مقایسه پاسخ آن با دزیمتر (LiF: Mg, Ti (TLD-100 ساخت شرکت هارشا در برابر پرتوهای گاما
Thermoluminescence dosimeters are small tablets used with 3. 2*3. 2*0. 9mm^3 for measurement of received dose from radioactive various beams. The most common dosimeter is TLD-100 which is lithium-fluoride family and this dosimeter contain magnesium impurities. In this study, first lithium-flouride powder was mixed with titanium and controlled atmosphere samples were heated. After measuring ...
full textquantitative evaluation of lif: mg, ti (tld-100) long-term sensitivity stability
background : sensitivity is an important dosimetric issue particularly in personal dosimetry where the range of doses is often very low to medium. in-service aging is assumed to be of negligible effect on the response of thermoluminescent dosimeters (tld), consequently, some service providers may mix fresh tlds to in-use ones and simply normalize the readings of tlds through batch calibration. ...
full textEvaluation of the Entrance Skin Dose in Animals Undergoing Diagnostic Radiology Using LiF, Mg, Ti Thermoluminescence Dosimetry (TLD-100)
Introduction According to the International Commission on Radiological Protection publication numbers 60 and 21, the environmental control standards should ensure human and other species protection to a desirable degree. Since application of radiographic procedures in Veterinary Medicine has increased significantly, in this study, we aimed to evaluate the entrance skin dose to the animals (e.g....
full textA Validation Study on Radiation properties of a Novel LiF: Mg, Ti Known as IAP-100
Introduction: LiF dosimeter has the most application in medicine. This study aimed to evaluate some dosimetric properties of a novel LiF: Mg, Ti. Materials and Methods: An ELEKTA Precise linear accelerator was used to calibrate dosimeters at 6 MV. In this survey, responses of dosimeters were evaluated up to 1000 cGy. Background effect was investigated in two different dosimeter states includin...
full textMy Resources
Journal title
volume 5 issue 4
pages -
publication date 2015-12-01
By following a journal you will be notified via email when a new issue of this journal is published.
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023