Investigation of Collimator Influential Parameter on SPECT Image Quality: a Monte Carlo Study

author

Abstract:

Background: Obtaining high quality images in Single Photon Emission Tomography (SPECT) device  is the most important goal in nuclear medicine. Because if image quality is low, the possibility of making a mistake in diagnosing and treating the patient will rise. Studying effective factors in spatial resolution of imaging systems is thus deemed to be vital. One of the most important factors in SPECT imaging in nuclear medicine is the use of an appropriate collimator for a certain radiopharmaceutical feature in order to create the best image as it can be effective in the quantity of Full Width at Half Maximum (FWHM) which is the main parameter  in spatial resolution.Method: In this research, the simulation of the detector and collimator of SPECT imaging device, Model HD3 made by Philips Co. and the investigation of important factors on the collimator were carried out using MCNP-4c code.Results: The results of the experimental measurments and simulation calculations revealed a relative difference of less than 5% leading to the confirmation of the accuracy of conducted simulation MCNP code calculation. Conclusion: This is the first essential step in the design and modelling of new collimators used for creating high quality images in nuclear medicine. 

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

investigation of collimator influential parameter on spect image quality: a monte carlo study

background: obtaining high quality images in single photon emission tomography (spect) device  is the most important goal in nuclear medicine. because if image quality is low, the possibility of making a mistake in diagnosing and treating the patient will rise. studying effective factors in spatial resolution of imaging systems is thus deemed to be vital. one of the most important factors in sp...

full text

Evaluation of the performance of parallel-hole collimator for high resolution small animal SPECT: A Monte Carlo study

Introduction: Image quality and accuracy of in vivo activity quantification in SPECT are affected by collimator penetration and scatter components, especially in high energy imaging. These phenomena highly depend on the collimator characteristic and photon energy. The presence of penetrated and scattered photons from collimator in SPECT images degrades spatial resolution, contr...

full text

Attenuation Correction in SPECT during Image Reconstruction using an Inverse Monte Carlo Method: A Simulation Study

Introduction: The main goal of SPECT imaging is to determine activity distribution inside the organs of the body. However, due to photon attenuation, it is almost impossible to do a quantitative study. In this paper, we suggest a mathematical relationship between activity distribution and its corresponding projections using a transfer matrix. Monte Carlo simulation was used to find a precise tr...

full text

SPECT Uptake Ratios: A Monte Carlo Study

In dopamine brain SPECT, semiquantitative techniques are in use, mostly for research purposes, to calculate activity uptake in the striatum relative to the background. The measured uptake ratios depend on both acquisition and reconstruction, and one important parameter is the rotation radius of the g-camera detectors, which affects spatial resolution. In brain SPECT research studies, the rotati...

full text

A Monte Carlo study of SPECT in boron neutron capture therapy for a heterogeneous human phantom

Background: Boron neutron capture therapy (BNCT) is a binary radiotherapy combining biochemical targeting with neutron irradiation. However, monitoring the boron distribution is a fundamental problem in BNCT. Prompt gamma rays emitted by boron capture reaction can be used to address the issue. Materials and Methods: The general-purpose Monte Carlo toolkits Geant4 and MCNP were used for the simu...

full text

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 5  issue 1

pages  -

publication date 2015-03-01

By following a journal you will be notified via email when a new issue of this journal is published.

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023