Fabrication of New 3D Phantom for the measurement of Geometric Distortion in Magnetic Resonance Imaging System
Authors
Abstract:
Introduction: Geometric distortion, an important parameter in neurology and oncology. The current study aimed to design and construct a new three-dimensional (3D) phantom using a 3D printer in order to measure geometric distortion and its 3D reproducibility. Material and Methods: In this study, a new phantom containing 13,824 reference features (control points) was designed with AutoCAD software, fabricated with a 3D printer, and filled with vegetable oil. This phantom was tested on the Siemens 3 Tesla Prisma MRI model using a 64-channel head coil. Six-slice computed tomography (CT) scan images were used as a reference. Moreover, the reference features of MRI images were matched with those of CT scan images using a 3D reference model. The reproducibility of the phantom was investigated on three different days (three different imaging sessions per day). Results: The obtained 3D results indicated that the non-uniformity of field and nonlinearity of the gradients and imaging reproducibility could lead to geometric distortion. The mean Euclidean distance error for MRI volume was less than 1 mm. The maximum Euclidean error was 1.5 mm. Distortion in the whole volume was pronounced more specifically at the edges of the magnetic field. Conclusion: The results showed that the amount of distortion in the middle of the field was less than at its sides. This phantom can be used to check the distortion filters on the device. Furthermore, this phantom can be used to study geometric distortion in scenarios that require a small study volume, such as prostate studies.
similar resources
Fabrication of New 3D Phantom for Measuring Geometric Distortion in Magnetic Resonance Imaging System
Introduction: Geometric distortion is a major shortcoming of magnetic resonance imaging (MRI), which has an important influence on the accuracy of volumetric measurements, an important parameter in neurology and oncology. Our goal is to design and construct a new three- dimensional phantom using a 3D printer in order to measure geometric distortion and its reproducibility in...
full textGeometric distortion evaluation of magnetic resonance images by a new large field of view phantom for magnetic resonance based radiotherapy purposes
Background: The magnetic resonance imaging (MRI)-based radiotherapy planning method have been considered in recent years because of the advantages of MRI and the problems of planning with two images modality. The first step in MRI-based radiotherapy is to evaluate magnetic resonance (MR) images geometric distortion. Therefore, the present study aimed to evaluate system related geometric distort...
full textDesign and Fabrication of a Four-Dimensional Respiratory Phantom for Studying Tumor Movement in Radiotherapy with Magnetic Resonance Imaging
Introduction: In radiation therapy, determining the location of the tumor accurately during irradiation is one of the most important requirements. However, lung tumors are not fixed in a single location and move during irradiation due to respiratory motion. Due to limitations in assessing such movements, using a lung phantom can be useful and operational for their fast, easy an...
full textGeometric Distortion in Structural Magnetic Resonance Imaging
Geometric distortion, an undesirable image artifact, is an inferior aspect associated with magnetic resonance imaging (MRI). Although slight distortions in MR images often have no consequences in reaching clinical conclusions, geometric distortions can make significant differences in certain MR applications such as, for example, stereotactic localization in radio-surgery and MR image-guided bio...
full textMeasurement of Posterior Tibial Slope Using Magnetic Resonance Imaging
Background: Posterior tibial slope (PTS) is an important factor in the knee joint biomechanics and one of the bone features, which leads to knee joint stability. Posterior tibial slope affects flexion gap, knee joint stability and posterior femoral rollback that are related to wide range of knee motion. During high tibial osteotomy and total knee arthroplasty (TKA) surgery, proper retaining the...
full textMy Resources
Journal title
volume 16 issue 5
pages 377- 384
publication date 2019-09-01
By following a journal you will be notified via email when a new issue of this journal is published.
Keywords
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023