Evaluation of the RtDosePlan Treatment Planning System using Radiochromic Film and Monte Carlo Simulation
Authors
Abstract:
Introduction: GafChromic EBT films are one of the self-developing and modern films commercially available for dosimetric verification of treatment planning systems (TPSs). Their high spatial resolution, low energy dependence and near-tissue equivalence make them suitable for verification of dose distributions in radiation therapy. This study was designed to evaluate the dosimetric parameters of the RtDosePlan TPS such as PDD curves, lateral beam profiles, and isodose curves measured in a water phantom using EBT Radiochromic film and EGSnrc Monte Carlo (MC) simulation. Methods and Materials: A Microtek color scanner was used as the film scanning system, where the response in the red color channel was extracted and used for the analyses. A calibration curve was measured using pieces of irradiated films to specific doses. The film was mounted inside the phantom parallel to the beam's central axis and was irradiated in a standard setup (SSD = 80 cm, FS = 10×10 cm2) with a 60Co machine. The BEAMnrc and the DOSXYZnrc codes were used to simulate the Co-60 machine and extracting the voxel-based phantom. The phantom's acquired CT data were transferred to the TPS using DICOM files. Results: Distance-To-Agreement (DTA) and Dose Difference (DD) among the TPS predictions, measurements and MC calculations were all within the acceptance criteria (DD=3%, DTA=3 mm). Conclusion: This study shows that EBT film is an appropriate tool for verification of 2D dose distribution predicted by a TPS system. In addition, it is concluded that MC simulation with the BEAMnrc code has a suitable potential for verifying dose distributions.
similar resources
evaluation of the rtdoseplan treatment planning system using radiochromic film and monte carlo simulation
introduction: gafchromic ebt films are one of the self-developing and modern films commercially available for dosimetric verification of treatment planning systems (tpss). their high spatial resolution, low energy dependence and near-tissue equivalence make them suitable for verification of dose distributions in radiation therapy. this study was designed to evaluate the dosimetric parameters of...
full textComparison of Electron-Beam Dose Distributions in a Heterogeneous Phantom Obtained Using Radiochromic Film Dosimetry and Monte Carlo Simulation
Introduction: Nowadays new radiochromic films have an essential role in radiotherapy dosimetry. Properties such as high sensitivity, good reproducibility, high spatial resolution, easy readout and portability have made them attractive for dosimetry, especially in high-dose-gradient regions. Material and Methods: In this study, electron-beam dose distributions in homogenous and heterogeneous pha...
full textDose distribution verification for GZP6 sources: a comparison of Monte Carlo, radiochromic film, and GZP6 treatment planning system
www.onk.ns.ac.rs/Archive Vol 20, No. 1–2, July 2012 INTRODUCTION The GZP6 high dose rate (HDR) afterloading brachytherapy unit (Nuclear Power Institute of China) (1) has recently been introduced in the Iranian radiation oncology centers. The unit has six different cobalt-60 source braids being loaded in tandem and ovoid applicators. The sources are used in the treatment of cervix, vagina, rectu...
full textMonte Carlo Simulation for Treatment Planning Optimization of the COMS and USC Eye Plaques Using the MCNP4C Code
Introduction: Ophthalmic plaque radiotherapy using I-125 radioactive seeds in removable episcleral plaques is often used in management of ophthalmic tumors. Radioactive seeds are fixed in a gold bowl-shaped plaque and the plaque is sutured to the scleral surface corresponding to the base of the intraocular tumor. This treatment allows for a localized radiation dose delivery to the tumor with a ...
full textCT-Based Brachytherapy Treatment Planning using Monte Carlo Simulation Aided by an Interface Software
Introduction: In brachytherapy, radioactive sources are placed close to the tumor, therefore, small changes in their positions can cause large changes in the dose distribution. This emphasizes the need for computerized treatment planning. The usual method for treatment planning of cervix brachytherapy uses conventional radiographs in the Manchester system. Nowadays, because of their advantages ...
full textcomparison of electron-beam dose distributions in a heterogeneous phantom obtained using radiochromic film dosimetry and monte carlo simulation
introduction: nowadays new radiochromic films have an essential role in radiotherapy dosimetry. properties such as high sensitivity, good reproducibility, high spatial resolution, easy readout and portability have made them attractive for dosimetry, especially in high-dose-gradient regions. material and methods: in this study, electron-beam dose distributions in homogenous and heterogeneous pha...
full textMy Resources
Journal title
volume 7 issue 2
pages 81- 93
publication date 2010-06-01