Estimation of Dosimetric Parameters based on KNR and KNCSF Correction Factors for Small Field Radiation Therapy at 6 and 18 MV Linac Energies using Monte Carlo Simulation Methods

Authors

Abstract:

Background: Estimating dosimetric parameters for small fields under non-reference conditions leads to significant errors if done based on conventional protocols used for large fields in reference conditions. Hence, further correction factors have been introduced to take into account the influence of spectral quality changes when various detectors are used in non-reference conditions at different depths and field sizes.Objective: Determining correction factors (KNR and KNCSF) recommended recently for small field dosimetry formalism by American Association of Physicists in Medicine (AAPM) for different detectors at 6 and 18 MV photon beams.Methods: EGSnrc Monte Carlo code was used to calculate the doses measured with different detectors located in a slab phantom and the recommended KNR and KNCSF correction factors for various circular small field sizes ranging from 5-30 mm diameters. KNR and KNCSF correction factors were determined for different active detectors (a pinpoint chamber, EDP-20 and EDP-10 diodes) in a homogeneous phantom irradiated to 6 and 18 MV photon beams of a Varian linac (2100C/D).Results: KNR correction factor estimated for the highest small circular field size of 30 mm diameter for the pinpoint chamber, EDP-20 and EDP-10 diodes were 0.993, 1.020 and 1.054; and 0.992, 1.054 and 1.005 for the 6 and 18 MV beams, respectively. The KNCSF correction factor estimated for the lowest circular field size of 5 mm for the pinpoint chamber, EDP-20 and EDP-10 diodes were 0.994, 1.023, and 1.040; and 1.000, 1.014, and 1.022 for the 6 and 18 MV photon beams, respectively.Conclusion: Comparing the results obtained for the detectors used in this study reveals that the unshielded diodes (EDP-20 and EDP-10) can confidently be recommended for small field dosimetry as their correction factors (KNR and KNCSF) was close to 1.0 for all small field sizes investigated and are mainly independent from the electron beam spot size.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

estimation of dosimetric parameters based on knr and kncsf correction factors for small field radiation therapy at 6 and 18 mv linac energies using monte carlo simulation methods

background: estimating dosimetric parameters for small fields under non-reference conditions leads to significant errors if done based on conventional protocols used for large fields in reference conditions. hence, further correction factors have been introduced to take into account the influence of spectral quality changes when various detectors are used in non-reference conditions at differen...

full text

Monte Carlo Simulation of Siemens Primus plus Linac for 6 and 18 MV Photon Beams

Objective: The aim of the present study is to simulate 6 MV and 18 MV photon beam energies of a Siemens Primus Plus medical linear accelerator (Linac) and to verify the simulation by comparing the results with the measured data.Methods: The main components of the head of Siemens Primus Plus linac were simulated using MCNPX Monte Carlo (MC) code. To verify the results, experimental data of perce...

full text

برآورد ضرایب تصحیح برای پرتودرمانی میدان‌های کوچک با استفاده از انرژی‌های ۶ و ۱۸ مگاولتی یک شتاب‌دهنده خطی

Abstract Background and purpose: Estimating dosimetric parameters for small fields at non-reference conditions lead to significant errors if they are done based on the conventional protocols used for large fields and reference conditions. The aim of this study was to determine and compare small fields correction factors (KNR and KNCSF) based on a novel protocol by using different types of d...

full text

Monte Carlo Simulation of Siemens Primus plus Linac for 6 and 18 MV Photon Beams

Objective The aim of the present study is to simulate 6 MV and 18 MV photon beam energies of a Siemens Primus Plus medical linear accelerator (Linac) and to verify the simulation by comparing the results with the measured data. Methods The main components of the head of Siemens Primus Plus linac were simulated using MCNPX Monte Carlo (MC) code. To verify the results, experimental data of perc...

full text

Monte carlo simulation of varian clinac iX 10 MV photon beam for small field dosimetry

Background: The lack of lateral electronic disequilibrium (LED) becomes a main problem in small field. This factor affects the dose in target volume cannot predict correctly. In addition, utilization of high-energy linear accelerator (10 MV) can emit some unwanted particles (electron contamination). Therefore, the aim of this study was to characterize head linear accelerator (linac) Varian Clin...

full text

A Monte Carlo simulation and dosimetric verification of physical wedges used in radiation therapy

Background: The presence of a wedge filter in the beam trajectory can modify the beam quality and cause some changes in the dosimetry parameters which are usually difficult to be measured directly and accurately. Material and Methods: In this study the MCNP-4C Monte Carlo code was used to simulate the 9 MV photon beam generated by a linear accelerator. Upon getting a good agreement bet...

full text

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 9  issue 1

pages  -

publication date 2019-02-01

By following a journal you will be notified via email when a new issue of this journal is published.

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023