Dose Reduction to the Thyroid Gland in Pediatric Chest Radiography

Authors

  • Mansour Zabihzadeh Assistant Professor, Department of Medical Physics, School of Medicine, Student Research Committee, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.
  • Mehrdad Gholami Assistant Professor, Department of Medical Physics, Lorestan University of Medical Sciences, Khorram Abad, Iran.
  • Nasim Shams Assistant Professor (PhD), Department of Oral and Maxillofacial radiology, School of dentistry, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran
  • Vahid Karami Medical Physics Student (MSc), Department of Medical Physics, Student Research Committee, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.
  • Zahra Fazeli Nezhad Radiologist, Department of Radiology, School of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.
Abstract:

Background  It is remain a main concern that pediatric chest radiographies contribute to the significant radiation exposure to the thyroid gland as a more susceptible organ to radiation induced cancer. The aim of this study was to evaluate the entrance surface dose (ESD) of pediatric chest radiography compared to the diagnostic reference levels (DRL) and evaluation the efficacy of the lead (Pb) shield in radiation dose reduction to the thyroid gland.Materials and Methods After assessing each patient against specific inclusion-exclusion criteria, 40 pediatric patients who were undergoing anterior-posterior (AP) projection of the chest x-ray were considered eligible for this study. The ESD of the chest and also ESD of thyroid gland with and without a 1 mm butterfly-shaped lead shield which placed on the thyroid gland were measured using high sensitive thermo luminescent dosimeters (TLD-GR 200).Results The average of ESD for chest radiography was 0.068+ 0.006 mGy (0.021 - 0.232 mGy). The unshielded average thyroid ESD was 0.065 + 0.003 mGy compared to the shielded average thyroid ESD of 0.001 + 0.0005 mGy. The use of Pb-shield produced a statistically significant decrease in the average thyroid dose by about 97% (P< 0.001). Conclusion The use of Pb-thyroid shield in the AP projection of pediatric chest radiography has potential to reduced radiation dose without compromising image quality.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

dose reduction to the thyroid gland in pediatric chest radiography

background  it is remain a main concern that pediatric chest radiographies contribute to the significant radiation exposure to the thyroid gland as a more susceptible organ to radiation induced cancer. the aim of this study was to evaluate the entrance surface dose (esd) of pediatric chest radiography compared to the diagnostic reference levels (drl) and evaluation the efficacy of the lead (pb)...

full text

An Estimate of Radiation Dose to the Lens of the Eyes, Parotid Gland, and Thyroid Gland in Dental Panoramic Radiography

Introduction: Dental panoramic radiography (DPR) is one of the most frequent diagnostic X-ray procedures, the application of which is currently on a growing trend. During DPR, several radiosensitive tissues, such as the lens of the eyes, parotid gland, and thyroid gland, contribute to the radiation field, and it is necessary to monitor their received dose. The aim of t...

full text

Efficacy of Increasing Focus to Film Distance (FFD) for Patient’s Dose and Image Quality in Pediatric Chest Radiography

Background: Increasing the x-ray focus to film distance (FFD) has been advocated as an effective method to reduce the patients’ radiation dose. The aim of this study was to investigate the efficacy of this technique in patients' dose and image quality in pediatric chest x-ray.Material and methods: Sixty pediatric patients were x-ray imaged at FFDs of 100 and 130 cm. Dose measurements were perfo...

full text

Dose Reduction Methods in Chest Computed Tomography

Introduction: Computed tomography is one of the most powerful tools for investigating thorax disease, because it shows progression of lung disease much clearer than chest radiography. Technical advances in CT scanners have led to some abilities for CT scanners such as coverage of larger volume, lower noise and ability for acquiring image in one respiratory phase. However, this will be accompani...

full text

Evaluation of the Entrance Surface Dose (ESD) and Radiation Dose to the Radiosensitive Organs in Pediatric Pelvic Radiography

Background Patients' dosimetry is crucial in order to enhance radiation protection optimization and to deliver low radiation dose to the patients in a radiological procedure. The aim of this study was to assess the entrance surface dose (ESD) and radiation dose to the radiosensitive organs in pediatric pelvic radiography. Materials and Methods The studied population included 98 pediatric patien...

full text

Computed Radiography dose optimization in pediatric patients

Radiation dose reduction in pediatric X-ray imaging is especially important because of children radiation sensitivity. For any radiographic examination performed at a fixed radiographic tube potential, the patient absorbed dose is directly proportional to the value of milliampere-seconds (mAs) selected by the operator. Nevertheless, reducing X-ray exposure has the unavoidable disadvantage of in...

full text

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 4  issue 7

pages  2183- 2191

publication date 2016-07-01

By following a journal you will be notified via email when a new issue of this journal is published.

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023