Abstract:
Background: Monte Carlo (MC) modeling of a linear accelerator is a prerequisite for Monte Carlo dose calculations in external beam radiotherapy. In this study, a simple and efficient model was developed for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code Materials and methods: The head of Elekta SL-25 linac was simulated for 6 and 18 MV photon beams using MCNP4C MC code. Energy spectra and fluence distribution of photons crossing the phase space plane were calculated. A simple point source model was developed based on calculated photon spectra and spatial distribution. Using this model, percent depth doses (PDDs) and beam profiles were calculated for different field sizes. The results of MC calculations were compared with measurements. Results: There was a good agreement between MC calculations and measurement for descending part of PDD curves. But, comparing calculated PDDs with measurement showed up to 10% differences for build up region of PDD curves for both energies. For beam profiles, there was 2% difference in flat region and up to 15% difference was seen for out of field region. These results were acceptable according to recommended criteria. Although, using our model the run time was decreased 24 times in comparison to original full Monte Carlo method. Conclusion: Our study showed that our model is accurate and effective for MC calculations in radiotherapy treatment planning. Also, it substantially lowers MC runtime for radiotherapy purposes.
similar resources
developement a simple point source model for elekta sl-25 linear accelerator using mcnp4c monte carlo code
background: monte carlo (mc) modeling of a linear accelerator is a prerequisite for monte carlo dose calculations in external beam radiotherapy. in this study, a simple and efficient model was developed for elekta sl-25 linear accelerator using mcnp4c monte carlo code materials and methods: the head of elekta sl-25 linac was simulated for 6 and 18 mv photon beams using mcnp4c mc code. energy sp...
full textA Monte Carlo study on Photoneutron Spectrum around Elekta SL75/25 18 MV linear accelerator
Medical linear accelerators are one of the most widespread methods for cancer treatment. Despite their advantages, unwanted photoneutrons are produced by high energy linacs. This photoneutrons are as undesired doses to patients and a significant problem for radiation protection of the staffs and patients. Photoneutrons radiological risk must be evaluated because of their high LET and range.in o...
full textA Monte Carlo study on Photoneutron Spectrum around Elekta SL75/25 18 MV linear accelerator
Medical linear accelerators are one of the most widespread methods for cancer treatment. Despite their advantages, unwanted photoneutrons are produced by high energy linacs. This photoneutrons are as undesired doses to patients and a significant problem for radiation protection of the staffs and patients. Photoneutrons radiological risk must be evaluated because of their high LET and range.in o...
full texta monte carlo study on photoneutron spectrum around elekta sl75/25 18 mv linear accelerator
medical linear accelerators are one of the most widespread methods for cancer treatment. despite their advantages, unwanted photoneutrons are produced by high energy linacs. this photoneutrons are as undesired doses to patients and a significant problem for radiation protection of the staffs and patients. photoneutrons radiological risk must be evaluated because of their high let and range.in o...
full textA comparative Monte Carlo study on 6MV photon beam characteristics of Varian 21EX and Elekta SL-25 linacs
Background: Monte Carlo method (MC) has played an important role in design and optimization of medical linacs head and beam modeling. The purpose of this study was to compare photon beam features of two commercial linacs, Varian 21EX and Elekta SL-25 using MCNP4C MC code. Materials and Methods: The 6MV photon beams of Varian 21EX and Elekta Sl-25 linacs were simulated based on manufacturers pro...