Determination of Dosimetric characteristics of a New 192Ir-PDR Brachytherapy Source According to AAPM TG- 43 Protocol using Monte Carlo simulation technique
Authors
Abstract:
Introduction: 192Ir is one of the important sources frequently used in brachytherapy. Up to now, a lot of commercial models of this source have been made which Ir-192 has been recently added to them. The aim of the present study is to determine the dosimetric parameters of this new source model based on the recommendations of TG-43(U1) protocol using Monte Carlo simulation technique. Materials and Methods: The assessed parameters included dose rate constant, radial dose function, anisotropy function, and dose distribution around the source. Calculations were performed in water phantom using MCNP4C Monte Carlo code. Results: The dose rate constant of this new source in water was 1.07+0.06 cGyh-1U-1 which was in a good agreement with the TLD measured data (1.13+0.07 cGyh- 1U-1). Conclusion: The results of this study showed that the dosimetric characteristics of this new brachytherapy source are comparable with those of other commercially available sources. Furthermore, the simulated parameters were in accordance with the previously measured ones. This means that the Monte Carlo simulated dosimetric parameters could be employed to obtain the dose distribution around this new brachytherapy source based on TG-43(U1) protocol.
similar resources
A segmented 32P source Monte Carlo model to derive AAPM TG-60 dosimetric parameters used for intravascular brachytherapy.
A new high dose rate 20 mm 32P intravascular brachytherapy (IVB) beta source used with automated stepping has recently been introduced. The AAPM Task Group 60 recommends that beta IVB sources should have well characterized dosimetric parameters in water. In this study, Monte Carlo simulations (MCNPX v 2.4) were used to derive these parameters for a 2 mm source segment rather than the entire 20 ...
full textA Comparison of the Dosimetric Parameters of Cs-137 Brachytherapy Source in Different Tissues with Water Using Monte Carlo Simulation
Introduction After the publication of Task Group number 43 dose calculation formalism by the American Association of Physicists in Medicine (AAPM), this method has been known as the most common dose calculation method in brachytherapy treatment planning. In this formalism, the water phantom is introduced as the reference dosimetry phantom, while the attenuation coefficient of the sources in the...
full textA Comparison of dosimetric parameters between IAEA TRS-398, AAPM TG-51 protocols and Monte-Carlo simulation
Background: Two protocols of AAPM TG-51 and IAEA TRS-398 were compared followed by a measurement and Monte Carlo simulation of beam quality correction factor, KQ, AAPM TG-51 and IAEA TRS-398 protocols were compared for the absorbed dose to water for DW, and KQ parameters. Materials and Methods: Dose measurements by either protocols were performed with cylindrical and plane parallel ch...
full textDetermination of Dosimetric Characteristics of IrSeed 125I Brachytherapy Source
Introduction Low dose rate brachytherapy sources have been widely used for interstitial implants in tumor sites, particularly in prostate. Dosimetric characteristics of a new IrSeed 125I brachytherapy source have been determined using the LiF thermoluminescent dosimeter (TLD) chips. Materials and Methods Dose rate constant, radial dose function, and anisotropy function around the IrSeed 125I so...
full textInvestigation of dose distribution 252Cf Isotron brachytherapy source based on TG-43U1 protocol by Monte Carlo method
Introduction:The commercial 252Cf sources are too large in size and clinical applications of neutronbrachytherapy (NBT) are limited to a small number of intracavitary treatments of cervical cancers. Recently, under the Cooperative Research and Development Agreement (CRADA) with Isotron Inc., the Oak Ridge National Laboratory (ORNL) encapsulated a new medical 252Cf sources, cal...
full textDosimetric Parameters Estimation of I-125 Brachytherapy Source in fat phantom using GATE8.0 code
Introduction: Brachytherapy is one type of internal radiation therapy where radiation sources, which are usually encapsulated, are placed as close as possible to the tumor site or inside the patient's body. In this technique, it is important to determine dose distribution around the brachytherapy capsule to create optimal treatment plant. In this way, dosimetric parameters are...
full textMy Resources
Journal title
volume 15 issue Special Issue-12th. Iranian Congress of Medical Physics
pages 402- 402
publication date 2018-12-01
By following a journal you will be notified via email when a new issue of this journal is published.
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023