Design and Construction of an Affordable Phantom for Electron Density Measurement and Linearity Tests of Computed Tomography Systems
Authors
Abstract:
Introduction: The performance of computed tomography is routinely checked using phantoms, which are known as important diagnostic imaging tools. Depending upon the aim of the study, different phantoms are designed, while trying to satisfy certain levels of diversities in their application. Material and Methods: The present study describes the construction of an inexpensive phantom designed for simultaneous measurements of 12 different samples. The body of the phantom, test tube holders, and test tubes were made of materials of low attenuation coefficient. Body of the phantom was filled with water. Test tubes filled with solutions of known chemical compositions were mounted on the test tube holders. The whole phantom was scanned at 80, 100, 120, 140 kVp to evaluate the performance of the CT system. Using Hounsfield Unit (HU) data from these liquid samples of known electron density, the phantom was calibrated for electron density measurements. Results: The system's accuracy and reproducibility were verified by measuring the HU values for some known materials. According to the results obtained from the experimental datawith liquid samples, the accuracy of the water and noise was within ±3.2 HU and 0.6%, respectively. Moreover, the image uniformity error was less than ±2 HU, and CT system's linearity for calibration was estimated with 99.9% confidence. Conclusion: The present system gives satisfactory results with known samples and can be used with confidence for characterizing unknown materials.
similar resources
development and implementation of an optimized control strategy for induction machine in an electric vehicle
in the area of automotive engineering there is a tendency to more electrification of power train. in this work control of an induction machine for the application of electric vehicle is investigated. through the changing operating point of the machine, adapting the rotor magnetization current seems to be useful to increase the machines efficiency. in the literature there are many approaches wh...
15 صفحه اولDesign, Construction and Evaluation of an Anthropomorphic Head Phantom for Assessment of Image Distortion in Stereotactic Radiosurgery Planning Systems
Introduction: In recent years, the use of magnetic resonance (MR) images in radiation treatment planning has drawn considerable attention. However, although the extent of a tumor can be determined in great detail on MR images, the geometric accuracy of these images is limited by distortions stemming from the inhomogeneity of the static background magnetic field, the nonlineari...
full textconstruction and validation of the translation teacher competency test and the scale of students’ perceptions of translation teachers
the major purpose of this study was to develop the translation teacher competency test (ttct) and examine its construct and predictive validity. the present study was conducted in two phases: a qualitative phase as well as a quantitative phase. in the first phase of the study, the author attempted to find out the major areas of competency required for an academic translation teacher. the second...
nano-rods zno as an efficient catalyst for the synthesis of chromene phosphonates, direct amidation and formylation of amines
چکیده ندارد.
simulation and design of electronic processing circuit for restaurants e-procurement system
the poor orientation of the restaurants toward the information technology has yet many unsolved issues in regards to the customers. one of these problems which lead the appeal list of later, and have a negative impact on the prestige of the restaurant is the case when the later does not respond on time to the customers’ needs, and which causes their dissatisfaction. this issue is really sensiti...
15 صفحه اولapplication and construction of carbon paste modified electrodes developed for determination of metal ions in some real samples
ساخت الکترودهاِی اصلاح شده ِیکِی از چالشهاِی همِیشگِی در دانش شیمِی بوِیژه شیمِی تجزیه مِی باشد ،که با در نظر گرفتن سادگِی ساخت، کاربردی بودن و ارزان بودن روش مِی توان به باارزش بودن چنِین سنسورهاِی پِی برد.آنچه که در ادامه آورده شده به ساخت و کاربرد الکترودهاِی اصلاح شده با استفاده از نانو ذرات در اندازه گِیرِی ولتامترِی آهن وکادمِیم اشاره دارد. کار اول اختصاص دارد به ساخت الکترود خمِیر کربن اصلاح شده با لِیگاند داِ...
15 صفحه اولMy Resources
Journal title
volume 17 issue 1
pages 38- 47
publication date 2020-01-01
By following a journal you will be notified via email when a new issue of this journal is published.
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023