Assessment of patients’ entrance skin and effective dose in a mathematical human phantom for the most common interventional radiological examinations
Authors
Abstract:
Introduction: Nowadays, we are witnessing an exponential use of interventional radiology techniques in different communities. After CT, interventional techniques are the second factor of increasing patients’ doses in different societies. Measuring patient doses from aforementioned methods has been recommended by many radiation protection professional organizations such as ICRP and IAEA. Our aim was to measure/calculate the patients’ entrance skin and also necessary parameters required to estimate relevant effective doses for common diagnostic and therapeutic interventional radiology examinations. Materials and Methods: After reviewing and analyzing interventional radiology examination data in an imaging center of a public hospital over 6 months, five most commonly used examinations, including cholangiography, chemoembolization liver, uterine fibroids embolization, bile duct stenting and coronary angiography and brain embolization were selected for the dose measurements/calculations. For each examination, 50 patients were selected and their skin doses were measured using TLDs. Effective doses resulted from these examinations were also calculated by using the Monte Carlo based PCXMC software for an average human phantom. Results: The average entrance skin dose measured for the patients undergoing cholangiography, chemoembolization liver, uterine fibroids embolization, bile duct stenting and coronary angiography and brain embolization examinations were: 57, 141, 447, 241, 587 mGy and their estimated effective doses were 3.5,11, 31, 15.5, 10 mSv respectively. As expected, the patient doses in interventional radiology examinations varied, since the amount of doses depends on many factors including the complexity and difficulty of interventional procedures, operators’ experience and skill, and patients’ weight. Conclusion: In interventional radiology examinations a small field of view is used. Hence, appropriate voltage potentials and pulse durations must be used to reduce effective doses while achieving highest image qualities (resolution, contrast-to-noise ratio, and signal to noise ratio).
similar resources
Assessment of patient entrance skin dose (ESD) and effective dose (ED) for the most common interventional radiological exams at Mazandaran hospital
Introduction: Nowadays we are witnessing an exponential application of interventional radiology techniques in various communities as well as Iran. Radiological technique is regarded as the 2nd contributing factor in increasing the amount of public doses in various societies. So that related radiation protection organization has recommended measuring patient d...
full textMeasurement of Entrance Skin Dose and Effective Dose in The most Common Diagnostic radiology Examinations in Jiroft, Iran
Introduction: Medical uses of radiation have grown very rapidly over in the past two decades, medical uses represent the largest source of exposure to people, The most important methods in medical diagnosis are the useses of diagnostic radiology exams. Although the use of these methods is very beneficial for the treatment of patients, but ionizing radiation produces ion pairs, ...
full textMeasurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran
The knowledge of the radiation dose received by the patient during the radiological examination is essential to prevent risks of exposures. The aim of this work is to study patient doses for common diagnostic radiographic examinations in hospitals affiliated to Kashan University of Medical sciences, Iran. The results of this survey are compared with those published by some national and internat...
full textAssessment of Maximum Entrance Skin Dose of Patients Undergoing Cardiac Interventional Procedures and Its Correlation with Other Dosimetric Parameters
Introduction: In recent years, the number of complex coronary angiography (CA) is increasing rapidly. These procedures have a significant contribution to medical exposure to the general population. Exposure of patients to high doses of x-rays could cause deterministic effects on the skin. Therefore, the assessment of radiation doses of patients is of great importance. ...
full textA Phantom Study for the Optimization of Image Quality and Radiation Dose for Common Radiographic Examinations in Digital Radiography
Introduction: Phantom studies facilitate the implementation of radiation dose surveillance as a function of radiographic technical parameters for minimizing patient radiation dose. The evidence of such investigations can then be used to evaluate technical parameters used in the radiographic procedures to reduce radiation dose without compromising the image quality. Material and Methods: This ex...
full textAssessment of Patients’ Entrance Skin Dose from Diagnostic X-ray Examinations at Public Hospitals of Akwa Ibom State, Nigeria
Introduction High doses of ionizing radiation can lead to adverse health outcomes such as cancer induction in humans. Although the consequences are less evident at very low radiation doses, the associated risks are of societal importance. This study aimed at assessing entrance skin doses (ESDs) in patients undergoing selected diagnostic X-ray examinations at public hospitals of Akwa Ibom State,...
full textMy Resources
Journal title
volume 15 issue Special Issue-12th. Iranian Congress of Medical Physics
pages 95- 95
publication date 2018-12-01
By following a journal you will be notified via email when a new issue of this journal is published.
Hosted on Doprax cloud platform doprax.com