An Approach in Radiation Therapy Treatment Planning: A Fast, GPU-Based Monte Carlo Method
Authors
Abstract:
Introduction: An accurate and fast radiation dose calculation is essential for successful radiation radiotherapy. The aim of this study was to implement a new graphic processing unit (GPU) based radiation therapy treatment planning for accurate and fast dose calculation in radiotherapy centers. Materials and Methods: A program was written for parallel running based on GPU. The code validation was performed by EGSnrc/ DOSXYZnrc. Moreover, a semi-automatic, rotary, asymmetric phantom was designed and produced using a bone, the lung, and the soft tissue equivalent materials. All measurements were performed using a Mapcheck dosimeter. The accuracy of the code was validated using the experimental data, which was obtained from the anthropomorphic phantom as the gold standard. Results: The findings showed that, compared with those of DOSXYZnrc in the virtual phantom and for most of the voxels (>95%), <3% dose-difference or 3 mm distance-to-agreement (DTA) was found. Moreover, considering the anthropomorphic phantom, compared to the Mapcheck dose measurements, <5% dosedifference or 5 mm DTA was observed. Conclusion: Fast calculation speed and high accuracy of GPU-based Monte Carlo method in dose calculation may be useful in routine radiation therapy centers as the core and main component of a treatment planning verification system.
similar resources
Review of Fast Monte Carlo Codes for Dose Calculation in Radiation Therapy Treatment Planning
An important requirement in radiation therapy is a fast and accurate treatment planning system. This system, using computed tomography (CT) data, direction, and characteristics of the beam, calculates the dose at all points of the patient's volume. The two main factors in treatment planning system are accuracy and speed. According to these factors, various generations of treatment planning syst...
full textCT-Based Brachytherapy Treatment Planning using Monte Carlo Simulation Aided by an Interface Software
Introduction: In brachytherapy, radioactive sources are placed close to the tumor, therefore, small changes in their positions can cause large changes in the dose distribution. This emphasizes the need for computerized treatment planning. The usual method for treatment planning of cervix brachytherapy uses conventional radiographs in the Manchester system. Nowadays, because of their advantages ...
full textGPU-based fast Monte Carlo simulation for radiotherapy dose calculation.
Monte Carlo (MC) simulation is commonly considered to be the most accurate dose calculation method in radiotherapy. However, its efficiency still requires improvement for many routine clinical applications. In this paper, we present our recent progress toward the development of a graphics processing unit (GPU)-based MC dose calculation package, gDPM v2.0. It utilizes the parallel computation ab...
full textMonte Carlo-based inverse treatment planning.
A Monte Carlo based inverse treatment planning system (MCI) has been developed which combines arguably the most accurate dose calculation method (Monte Carlo particle transport) with a 'guaranteed' optimization method (simulated annealing). A distribution of photons is specified in the tumour volume; they are transported using an adjoint calculation method to outside the patient surface to buil...