Advantages and disadvantages of removed flattening filter Linac: A Monte Carlo study on full quantitative dosimetric data of 18 MV-Varian Linac

Authors

  • Mansour Zabihzadeh Cancer, Environmental and Petroleum Pollutants Research Center, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran. Associate Professor, Department of Medical Physics, Faculty of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran. Associate Professor, Department of Clinical Oncology, Faculty of Medicine, Golestan Hospital, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.
  • Mohammad Reza Ay Professor, Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran. Research Center for Molecular and Cellular Imaging, Tehran University of Medical Sciences, Tehran, Iran
  • Mojtaba Hoseini- Ghahfarokhi PhD student, Department of Medical Physics, Faculty of Medicine, Ahvaz Jundishapur University of Medical Sciences, Ahvaz, Iran.
  • Seyed Rabee Mahdavi Associate Professor, Department of Medical Physics, Iran University of Medical Sciences, Tehran, Iran.
Abstract:

Introduction: During intensity modulated radiation therapy (IMRT) technique, theoretically, presence of flattening filter (FF) across the beamline of clinical linear accelerator (linac) is not essential to obtain uniform lateral profiles due to intensity modulation of photon beams by multileaf collimators (MLCs). The aim of this study was to investigate the dosimetrical properties of 18 MV photon Beam-Varian linac with and without FF. Materials and Methods: All dose measurements were performed on 18 MV, FF Mode-Varian 2100C/D linac. The FF and flattening filter free (FFF) modes of linac were modeled by MCNPX 2.4. code. The photon and contaminant electrons spectra, dose rate, present depth doses (PDD), lateral dose profiles, total and collimator scatter factors and out of field doses were calculated and compared with and without FF.   Results: Removing the FF increased the photon and contaminant electron fluences by factors of 5.48 and 3.94 for a 5 × 5 cm2 field size, respectively. The surface dose increased up to 155%. The flatness of dose profile was disturbed and deteriorated with increase of field size. Despite the dependence of the total scattering factor on field size, the variation of collimator scattering factors was neglected. The out-of-field dose decreased about 81.5 % for a 5 × 5 cm2 field size.   Conclusion: Removing FF from the linac’s head increases the dose rate and decreases the out-of-field dose, but the increased skin dose and deteriorated flatness of lateral dose profile are the main

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

Monte Carlo Simulation of Siemens Primus plus Linac for 6 and 18 MV Photon Beams

Objective: The aim of the present study is to simulate 6 MV and 18 MV photon beam energies of a Siemens Primus Plus medical linear accelerator (Linac) and to verify the simulation by comparing the results with the measured data.Methods: The main components of the head of Siemens Primus Plus linac were simulated using MCNPX Monte Carlo (MC) code. To verify the results, experimental data of perce...

full text

Enhancement of the Dose on 12 MV Linac with Free Flattening Filter Mode

Purpose: In the last years, some studies investigated dosimetric benefits of a free flattening filter for the photon mode in the radiotherapy field. This study aims to provide a theoretical study published and analysis of basic dosimetric properties for a Saturne 43 Linac head to implement free flattening filter beams clinically. Material and Methods: Thi...

full text

Estimation of Dosimetric Parameters based on KNR and KNCSF Correction Factors for Small Field Radiation Therapy at 6 and 18 MV Linac Energies using Monte Carlo Simulation Methods

Background: Estimating dosimetric parameters for small fields under non-reference conditions leads to significant errors if done based on conventional protocols used for large fields in reference conditions. Hence, further correction factors have been introduced to take into account the influence of spectral quality changes when various detectors are used in non-reference conditions at differen...

full text

A full quantitative analysis of 18 MV photon beam from 2100 C/D-Varian clinical linear accelerator with and without flattening filter

Background: During intensity modulated radiation therapy (IMRT) technique, theoretically, presence of flattening filter (FF) across the beamline of clinical linear accelerator (linac) is not essential to obtain uniform lateral profiles due to intensity modulation of photon beams by multileaf collimators (MLCs). The aim of this study was to investigate the dosimetrical properties of 18 MV photon...

full text

The plastic tray effect on the spectrum of photoneutrons prudused by 18 MV medical linac using the Monte Carlo method

Introduction: One of the important problems of using high energy linacs is the production of photoneutrons. High energy medical linacs produce secondary neutrons. These phtoneutrons increase patient dose   Materials and Methods: In this study, the effect of plastic trays on spectrum of photoneutrons produced by 18MV linac is investigated using the MCNP4C code...

full text

Monte Carlo Simulation of Siemens Primus plus Linac for 6 and 18 MV Photon Beams

Objective The aim of the present study is to simulate 6 MV and 18 MV photon beam energies of a Siemens Primus Plus medical linear accelerator (Linac) and to verify the simulation by comparing the results with the measured data. Methods The main components of the head of Siemens Primus Plus linac were simulated using MCNPX Monte Carlo (MC) code. To verify the results, experimental data of perc...

full text

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 15  issue Special Issue-12th. Iranian Congress of Medical Physics

pages  446- 446

publication date 2018-12-01

By following a journal you will be notified via email when a new issue of this journal is published.

Hosted on Doprax cloud platform doprax.com