Accuracy Evaluation of Oncentraâ„¢ TPS in HDR Brachytherapy of Nasopharynx Cancer Using EGSnrc Monte Carlo Code
Authors
Abstract:
Background: HDR brachytherapy is one of the commonest methods of nasopharyngeal cancer treatment. In this method, depending on how advanced one tumor is, 2 to 6 Gy dose as intracavitary brachytherapy is prescribed. Due to high dose rate and tumor location, accuracy evaluation of treatment planning system (TPS) is particularly important. Common methods used in TPS dosimetry are based on computations in a homogeneous phantom. Heterogeneous phantoms, especially patient-specific voxel phantoms can increase dosimetric accuracy.Materials and Methods: In this study, using CT images taken from a patient and ctcreate-which is a part of the DOSXYZnrc computational code, patient-specific phantom was made. Dose distribution was plotted by DOSXYZnrc and compared with TPS one. Also, by extracting the voxels absorbed dose in treatment volume, dose-volume histograms (DVH) was plotted and compared with Oncentraâ„¢ TPS DVHs.Results: The results from calculations were compared with data from Oncentraâ„¢ treatment planning system and it was observed that TPS calculation predicts lower dose in areas near the source, and higher dose in areas far from the source relative to MC code. Absorbed dose values in the voxels also showed that TPS reports D90 value is 40% higher than the Monte Carlo method.Conclusion: Today, most treatment planning systems use TG-43 protocol. This protocol may results in errors such as neglecting tissue heterogeneity, scattered radiation as well as applicator attenuation. Due to these errors, AAPM emphasized departing from TG-43 protocol and approaching new brachytherapy protocol TG-186 in which patient-specific phantom is used and heterogeneities are affected in dosimetry.Â
similar resources
Accuracy Evaluation of Oncentra™ TPS in HDR Brachytherapy of Nasopharynx Cancer Using EGSnrc Monte Carlo Code
BACKGROUND HDR brachytherapy is one of the commonest methods of nasopharyngeal cancer treatment. In this method, depending on how advanced one tumor is, 2 to 6 Gy dose as intracavitary brachytherapy is prescribed. Due to high dose rate and tumor location, accuracy evaluation of treatment planning system (TPS) is particularly important. Common methods used in TPS dosimetry are based on computati...
full textaccuracy evaluation of oncentra™ tps in hdr brachytherapy of nasopharynx cancer using egsnrc monte carlo code
background: hdr brachytherapy is one of the commonest methods of nasopharyngeal cancer treatment. in this method, depending on how advanced one tumor is, 2 to 6 gy dose as intracavitary brachytherapy is prescribed. due to high dose rate and tumor location, accuracy evaluation of treatment planning system (tps) is particularly important. common methods used in tps dosimetry are based on computat...
full textEGSnrc Monte Carlo-aided dosimetric studies of the new BEBIG 60Co HDR brachytherapy source
PURPOSE The purpose of this study is to obtain the dosimetric parameters of the new BEBIG (60)Co brachytherapy source following by TG-43U1 recommendation with appropriate electron cutoff energy (0.521 MeV). MATERIAL AND METHODS The new BEBIG (60)Co brachytherapy source is used to calculate the TG-43U1 parameters. EGSnrc-based Monte Carlo simulation code has been used to calculate the radial d...
full textMonte Carlo investigation on precise dosimetry of HDR breast brachytherapy with Accuboost
Introduction: Accuboost is a HDR brachytherapy system in early stages breast cancer treatment. This device provides a completely noninvasive procedure with parallel-opposed radiation from two immobilizing peripheral applicators that caused it a preferred option of modalities to choose. In most commercial treatment planning systems, tissues are considered as a simple water phan...
full textEvaluation of Dose Distribution Accuracy in HDR Brachytherapy of Esophagus Cancer Based on MRI Normoxic Polymer Gel Dosimetry
Introduction: The purpose of this work was to study the ability of MRI normoxic polymer gel dosimetry for evaluating the dose distribution in HDR brachytherapy of esophagial cancer at Imam Reza brachytherapy center (Mashhad, Iran). Materials and Methods: Initially, 2liters of normoxic gel (MAGIC) was fabricated and then poured into 12 calibration test tubes and placed in a perspex walled phanto...
full textSiemens primus accelerator simulation using EGSnrc Monte Carlo code and gel dosimetry validation with optical computed tomography system by EGSnrc code
Monte Carlo method is the most accurate method for simulation of radiation therapy equipment. The linear accelerators (linac) are currently the most widely used machines in radiation therapy centers. Monte Carlo modeling of the Siemens Primus linear accelerator in 6 MeV beams was used. Square field size of 10 × 10 cm2 produced by the jaws was compared with TLD. Head simulation of Siemens accele...
full textMy Resources
Journal title
volume 5 issue 1
pages -
publication date 2015-03-01
By following a journal you will be notified via email when a new issue of this journal is published.
Keywords
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023