A novel phantom design for brachytherapy quality assurance
Authors
Abstract:
Background: One major challenge in brachytherapy is to verify the accuracy of dose distributions calculated by the treatment planning system. In this project, a new phantom design has been introduced for quality assurance of dose distributions in gynocological (GYN) brachytherapy implants using EBT GafChromic film. Materials and Methods: This phantom has been designed and fabricated from 90 slabs of 18×16×0.2 cm3 Perspex to accommodate a tandem and ovoids assembly, which is normally used for GYN brachytherapy treatment. In addition, this phantom design is allowing the use EBT GafChromic films for dosimetric verification of GYN implants with Cs-137 Selectron LDR system. With this assembly, GafChromic films were exposed using a plan designed to deliver 2.5 Gy dose to point "A" in Manchester system for tandem and ovoids configurations and to deliver 1.5 Gy of dose to 0.5 cm distance from the lateral surface of ovoids for using ovoid-pair. The measured dose distributions with GafChromic films were compared with the TPS isodose lines both numerically and spatially. For a quantitative analysis of the results, the measured doses values at several points of interest were evaluated with the treatment planning data and values obtained following the TG-43 dose calculation formalism. Results: The results of these investigations have indicated that the new phantom design enables us to measure differences of greater than ±6% for LDR brachytherapy GYN treatments. Conclusion: The new phantom design could be utilized for the QA procedure of the brachytherapy remote after loading systems to confirm the accuracy of dose distribution in GYN implants.
similar resources
A quality assurance phantom for electronic portal imaging devices
Electronic portal imaging device (EPID) plays an important role in radiation therapy portal imaging, geometric and dosimetric verification. Consistent image quality and stable radiation response is necessary for proper utilization that requires routine quality assurance (QA). A commercial 'EPID QC' phantom weighing 3.8 kg with a dimension of 25 × 25 × 4.8 cm³ is used for EPID QA. This device ha...
full textDeveloping a Verification and Training Phantom for Gynecological Brachytherapy System
Introduction Dosimetric accuracy is a major issue in the quality assurance (QA) program for treatment planning systems (TPS). An important contribution to this process has been a proper dosimetry method to guarantee the accuracy of delivered dose to the tumor. In brachytherapy (BT) of gynecological (Gyn) cancer it is usual to insert a combination of tandem and ovoid applicators with a complicat...
full textThe evaluation of a 2D diode array in "magic phantom" for use in high dose rate brachytherapy pretreatment quality assurance.
PURPOSE High dose rate (HDR) brachytherapy is a treatment method that is used increasingly worldwide. The development of a sound quality assurance program for the verification of treatment deliveries can be challenging due to the high source activity utilized and the need for precise measurements of dwell positions and times. This paper describes the application of a novel phantom, based on a 2...
full textA Specific Patient Quality Assurance (PSQA) procedure for a Co-60 source based High Dose Rate Brachytherapy
Introduction: In radiation therapy, accurate dose determination and precise dose delivery to the tumor are directly associated with better treatment outcomes in terms of higher tumor control and lower post radiation therapy complications. The current study aims the development and clinical application of the Patient Specific Quality Assurance (PSQA) procedures for nasopharyngea...
full textTeflon cylindrical phantom for delivery quality assurance of SBRT
At our institution the standard delivery quality assurance (DQA) procedure for tomotherapy plans is accomplished with a water equivalent phantom, EDR2 film, and ion chamber point-dose measurements. Most plans deliver at most 5 Gy to the dose plane; however, recently a stereotactic body radiotherapy (SBRT) protocol has produced plans delivering upwards of 12 Gy to the film plane. EDR2 film satur...
full textDaily quality assurance phantom for ultrasound image guided radiation therapy
A simple phantom was designed, constructed, tested, and clinically implemented for daily quality assurance (QA) of an ultrasound-image-guided radiation therapy (US-IGRT) system, the Restitu Ultrasound system (Resonant Medical, Montreal, QC). The phantom consists of a high signal echogenic background gel surrounding a low signal hypoechoic egg-shaped target. Daily QA checks involve ultrasound im...