A novel dual energy CT-based attenuation correction method in PET/CT systems: A phantom study
Authors
Abstract:
In present PET/CT scanners, PET attenuation correction is performed by relying on the information given by CT scan. In the CT-based attenuation correction methods, dual-energy technique (DECT) is the most accurate approach, which has been limited due to the increasing patient dose. In this feasibility study, we have introduced a new method that can implement dual-energy technique with only a single energy CT scan. The implementation was done by CT scans of RANDO phantom at tube voltages of 80 kVP and 140 kVP. The acquired data was used to obtain conversion curves (which scale CT numbers at different kVP to each other), in three regions including lung tissue (HU<-100), soft tissue (-100200) for the combination of 80 kVP /140 kVP. Therefore, with having the CT image in one energy, we generate the CT image at the second energy (from now we call it virtual dual-energy technique) using these kVP conversion curves. The attenuation map at 511 keV was generated using bilinear (the most commonly used method in commercially available PET/CT scanners), real dual-energy and virtual dual-energy technique in a polyethylene phantom. In the phantom study, the created attenuation map using mentioned methods are compared to the theoretical values calculated using XCOM cross section library. The results in the phantom data show 10.1 %, 4.2 % and 4.3 % errors for bilinear, dual-energy and virtual dual-energy techniques respectively. Further evaluation using a larger patient data is underway to evaluate the potential of the technique in a clinical setting.
similar resources
a novel dual energy ct-based attenuation correction method in pet/ct systems: a phantom study
in present pet/ct scanners, pet attenuation correction is performed by relying on the information given by ct scan. in the ct-based attenuation correction methods, dual-energy technique (dect) is the most accurate approach, which has been limited due to the increasing patient dose. in this feasibility study, we have introduced a new method that can implement dual-energy technique with only a si...
full textA Novel Dual Energy CT-Based Attenuation Correction Method in PET/CT Systems: A Phantom Study
In present PET/CT scanners, PET attenuation correction is performed by relying on the information given by CT scan. In the CT-based attenuation correction methods, dual-energy technique (DECT) is the most accurate approach, which has been limited due to the increasing patient dose. In this feasibility study, we have introduced a new method that can implement dual-energy technique with only a si...
full textA novel energy mapping approach for CT-based attenuation correction in PET.
PURPOSE Dual-energy CT (DECT) is arguably the most accurate energy mapping technique in CT-based attenuation correction (CTAC) implemented on hybrid PET/CT systems. However, this approach is not attractive for clinical use owing to increased patient dose. The authors propose a novel energy mapping approach referred to as virtual DECT (VDECT) taking advantage of the DECT formulation but using CT...
full textComparing 511 keV Attenuation Maps Obtained from Different Energy Mapping Methods for CT Based Attenuation Correction of PET Data
Introduction: The advent of dual-modality PET/CT scanners has revolutionized clinical oncology by improving lesion localization and facilitating treatment planning for radiotherapy. In addition, the use of CT images for CT-based attenuation correction (CTAC) decreases the overall scanning time and creates a noise-free attenuation map (6map). CTAC methods include scaling, s...
full textComparative Assessment of Different Energy Mapping Approaches in CT Based Attenuation Correction: a Patient Study
Attenuation correction of PET emission data using spatially correlated CT images is fast and precise yielding a noise-free attenuation map (μmap) in comparison with radionuclide transmission scanning (TX). However, it is essential to convert the linear attenuation coefficients obtained from CT scans to those corresponding to 511 keV. Several conversion strategies have been developed including s...
full textAttenuation Correction in SPECT during Image Reconstruction using an Inverse Monte Carlo Method: A Simulation Study
Introduction: The main goal of SPECT imaging is to determine activity distribution inside the organs of the body. However, due to photon attenuation, it is almost impossible to do a quantitative study. In this paper, we suggest a mathematical relationship between activity distribution and its corresponding projections using a transfer matrix. Monte Carlo simulation was used to find a precise tr...
full textMy Resources
Journal title
volume 17 issue 2
pages 42- 49
publication date 2009-12-01
By following a journal you will be notified via email when a new issue of this journal is published.
Keywords
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023