A Monte Carlo study on Photoneutron Spectrum around Elekta SL75/25 18 MV linear accelerator

Authors

  • Asra Sadat Talebi M.Sc Graduated of Medical Physics, Department of Medical Physics, Semnan University of Medical Sciences, Semnan, Iran
  • Majid Jadidi Associate Professor, Department of Medical Physics, Semnan University of Medical Sciences, Semnan, Iran
  • Payman Hejazi Assistant Professor, Department of Medical Physics, Semnan University of Medical Sciences, Semnan, Iran
  • Raheb Ghorbani Professor, Head of Social Determinants of Health Research Center, Semnan University of Medical Sciences, Semnan, Iran
Abstract:

Medical linear accelerators are one of the most widespread methods for cancer treatment. Despite their advantages, unwanted photoneutrons are produced by high energy linacs. This photoneutrons are as undesired doses to patients and a significant problem for radiation protection of the staffs and patients. Photoneutrons radiological risk must be evaluated because of their high LET and range.in order to achieving this aim, photoneutron spectrum are calculated. The head of linac and a common treatment room was simulated by the MC code of MCNPX. Photoneutron spectrum was calculated in different field sizes, distances from isocenter and different cases (with and without structures and materials such as flattening filter, compensator, air and treatment room walls).The inclusion of the flattening filter and compensator had not any effects on shaping the photoneutron spectrum but neutron fluence and the average neutron energy are reduced obviously. Also effect of air on photoneutron spectrum was negligible. The calculation of photoneutron spectrum with concrete walls show that the component of fast neutrons is decreased and thermal neutrons are increased due to the room-return. In this case, with increasing distance from isocenter, fast neutrons are decreased and thermal neutrons are increased. As the field size is increased from 5×5 to 15×15 cm2, the neutron flux is increased clearly in isocenter. The neutrons flux are decreased near the door due to maze effect. The photoneutron spectrum investigation and risk estimation due to inclusion of neutron contamination in treatment room prevent from secondary cancer mortality.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

a monte carlo study on photoneutron spectrum around elekta sl75/25 18 mv linear accelerator

medical linear accelerators are one of the most widespread methods for cancer treatment. despite their advantages, unwanted photoneutrons are produced by high energy linacs. this photoneutrons are as undesired doses to patients and a significant problem for radiation protection of the staffs and patients. photoneutrons radiological risk must be evaluated because of their high let and range.in o...

full text

A Monte Carlo Study for Photoneutron Dose Estimations around the High-Energy Linacs

Background: High-energy linear accelerator (linac) is a valuable tool and the most commonly device for external beam radiation treatments in cancer patients. In the linac head, high-energy photons with energies above the threshold of interaction produce photoneutrons. These photoneutrons deliver the extra dose to the patients in the planning treatment and increase the risk of secondary cancer.O...

full text

Developement a simple point source model for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code

Background: Monte Carlo (MC) modeling of a linear accelerator is a prerequisite for Monte Carlo dose calculations in external beam radiotherapy. In this study, a simple and efficient model was developed for Elekta SL-25 linear accelerator using MCNP4C Monte Carlo code Materials and methods: The head of Elekta SL-25 linac was simulated for 6 and 18 MV photon beams using MCNP4C MC code. Energ...

full text

Photoneutron intensity variation with field size around radiotherapy linear accelerator 18-MeV X-ray beam

In X-ray radiotherapy accelerators, neutrons are produced mainly by ( , n) reaction when high energy X-rays interact with high Z materials of the linear accelerator head. These materials include the lead (Pb) used as shielding in the collimator, tungsten (W) target used for the production of X-rays and iron (Fe) in the accelerator head. These unwanted neutrons contaminate the therapeutic beam a...

full text

A Monte Carlo Study for Photoneutron Dose Estimations around the High-Energy Linacs

BACKGROUND High-energy linear accelerator (linac) is a valuable tool and the most commonly used device for external beam radiation treatments in cancer patients. In the linac head, high-energy photons with energies above the threshold of (γ,n) interaction produce photoneutrons. These photoneutrons deliver the extra dose to the patients undergoing radiation treatment and increase the risk of sec...

full text

My Resources

Save resource for easier access later

Save to my library Already added to my library

{@ msg_add @}


Journal title

volume 4  issue 1

pages  48- 57

publication date 2016-03-01

By following a journal you will be notified via email when a new issue of this journal is published.

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023