A High-Performance Liquid Chromatographic Assay for the Determination of Losartan in Plasma
Authors
Abstract:
A rapid and sensitive HPLC method was developed for determination of losartan in plasma. Losartan was extracted from plasma by a two-step extraction procedure using chloroform as extracting solvent in acidic medium. HPLC analysis was performed on a cyano reversed-phase column using phosphate buffer (pH 4.3), acetonitrile (750:250, v/v) as mobile phase with a flow rate of 0.9 mL/min. Sodium diclofenac was selected as internal standard. Excellent linearity between the peak area ratios and losartan concentrations over the range of 2-200 ng/mL of plasma was observed. The limit of determination with UV detection at 225 nm, with a CV < 5% was 2 ng/mL in 500 L of plasma sample. The assay was rapid, safe and reliable for use in pharmacokinetic studies of losartan in human being.
similar resources
High-Performance Liquid Chromatographic Determination of Rivastig-mine in Human Plasma for Application in Pharmacokinetic Studies
A simple and reproducible HPLC method with spectrophotometric detection was developed for determination of rivastigmine in human plasma. Liquid-liquid extraction of rivastigmine and donepezil (as internal standard) from plasma samples was performed with 1-butanol/n-hexane (2:98 v/v) in alkaline condition followed by back-extraction into diluted acetic acid. Chromatography was carried out using ...
full textComparison of Microbiological and High-Performance Liquid Chromatographic Methods for Determination of Clarithromycin Levels in Plasma
An agar well diffusion bioassay method for determination of clarithromycin in human plasma, using Micrococcus Luteus ATCC 9341 as the assay organism, was compared with a selective high-performance liquid chromatographic (HPLC) method with UV detection. Spiked plasma was used to prepare standard and control samples for both methods. The results of the bioassay analyses with spiked plasma samples...
full textComparison of Microbiological and High-Performance Liquid Chromatographic Methods for Determination of Clarithromycin Levels in Plasma
An agar well diffusion bioassay method for determination of clarithromycin in human plasma, using Micrococcus Luteus ATCC 9341 as the assay organism, was compared with a selective high-performance liquid chromatographic (HPLC) method with UV detection. Spiked plasma was used to prepare standard and control samples for both methods. The results of the bioassay analyses with spiked plasma samples...
full textHigh-Performance Liquid Chromatographic Determination of Rivastig-mine in Human Plasma for Application in Pharmacokinetic Studies
A simple and reproducible HPLC method with spectrophotometric detection was developed for determination of rivastigmine in human plasma. Liquid-liquid extraction of rivastigmine and donepezil (as internal standard) from plasma samples was performed with 1-butanol/n-hexane (2:98 v/v) in alkaline condition followed by back-extraction into diluted acetic acid. Chromatography was carried out using ...
full textSimple High-Performance Liquid Chromatographic Method for Determination of Ciprofloxacin in Human Plasma
A rapid, simple and sensitive high-performance liquid chromatography method was developed for determination of ciprofloxacin in plasma by means of ultraviolet detection. Ofloxacin was used as an internal standard and separation carried on a Novapak C18 column using a mobile phase of 0.01 M phosphate buffer (pH =2.6): methanol (82: 18 v/v). Extraction of drug was performed from plasma by liquid-...
full textMy Resources
Journal title
volume 2 issue 1
pages 18- 0
publication date 2003-05
By following a journal you will be notified via email when a new issue of this journal is published.
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023