تولید دزیمتر LiF: Mg, Ti و مقایسه پاسخ آن با دزیمتر (LiF: Mg, Ti (TLD-100 ساخت شرکت هارشا در برابر پرتوهای گاما
Authors
Abstract:
Thermoluminescence dosimeters are small tablets used with 3. 2*3. 2*0. 9mm^3 for measurement of received dose from radioactive various beams. The most common dosimeter is TLD-100 which is lithium-fluoride family and this dosimeter contain magnesium impurities. In this study, first lithium-flouride powder was mixed with titanium and controlled atmosphere samples were heated. After measuring of samples density and hardness, their glow curves were drawn and microscopic pictures of producted samples were provided. Of course, the reader should know that the best press pressure, the best range of temperature heating and the best range of samples heating were determined in 6-8 (ton/cm2), 775-800 ° C and 15-20 h, respectively [1]. Then, the response of producted TLD-100 under radiation of gamma sources, 60Co, was measured in 500 mSv rate and obtained values were compared with obtained values by Harshaw producted samples. Dosimetry characteristics of producted samples was evaluated according to ASTM E 668-00, IEC-ISO 1066 standards. The results of comparisions indicate good agreements between producted sample and Harshaw sample.
similar resources
تولید دزیمتر lif: mg, ti و مقایسه پاسخ آن با دزیمتر (lif: mg, ti (tld-۱۰۰ ساخت شرکت هارشا در برابر پرتوهای گاما
دزیمترهای ترمولومینسانس، قرص های کوچکی هستند که برای اندازه گیری دز دریافتی ناشی از پرتوهای مختلف رادیواکتیو مورد استفاده قرار می گیرند. معمول ترین آنها دزیمتر tld-100 بوده که از خانواده فلوراید- لیتوم و حاوی ناخالصی های منیزیوم و تیتانیوم می باشد. در این تحقیق، ابتدا پودر فلوراید- لیتیوم با ناخالصی منیزیوم و تیتانیوم مخلوط شده و سپس پرس شدند و در ادامه نمونه ها به مدت زمان مشخص و تحت فشار کنتر...
full textInvestigation of LiF, Mg and Ti (TLD-100) Reproducibility
LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the values of Element Correction Coefficient (ECC) were obtained 4 times by irradiating them to doses...
full textInvestigation of LiF, Mg and Ti (TLD-100) Reproducibility
LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the values of Element Correction Coefficient (ECC) were obtained 4 times by irradiating them to doses...
full textinvestigation of lif, mg and ti (tld-100) reproducibility
lif, mg and ti cubical tld chips (known as tld-100) are widely used for dosimetry purposes. the repeatability of tl dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mgy) with 662kev photons of cs-137. a group of 40 cubical tld chips was randomly selected from a batch and the values of element correction coefficient (ecc) were obtained 4 times by irradiating them to doses...
full textرشد بلور Lif با ناخالصیهای Mg و Ti به منظور ساخت ماده اولیه دزیمتر شخصی گرما - لیانی
تک بلورهای LiF
full textتعیین پارامترهای فروکشی دمایی دزیمتر ترمولومینسانس (TLD-100) LiF:Mg,Ti
Thermal quenching is a known and important effect which should be taken into account in studying thermoluminescence (TL) phenomenon. This effect causes the probability of non-radiative recombinations to be increased with raising temperature in TL process. This article deals with changes of the TL glow curve of LiF:Mg, Ti (TLD-100) with heating rate. It was observed that the main dosimetry peak ...