effectiveness evaluation of skin covers against intravascular brachytherapy sources using varskin3 code
Authors
abstract
background and objective: the most common intravascular brachytherapy sources include 32p, 188re, 106rh and 90sr/90y. in this research, skin absorbed dose for different covering materials in dealing with these sources were evaluated and the best covering material for skin protection and reduction of absorbed dose by radiation staff was recognized and recommended. method: four materials including polyethylene, cotton and two different kinds of plastic were proposed as skin covers and skin absorbed dose at different depths for each kind of the materials was calculated separately using the varskin3 code. results: the results showed that for all sources, skin absorbed dose was minimized when using polyethylene. considering this material as skin cover, maximum and minimum doses at skin surface were related to 90sr/90y and 106rh, respectively. conclusion: polyethylene was found the most effective cover in reducing skin dose and protecting the skin. furthermore, proper agreement between the results of varskin3 and other experimental measurements indicated that vraskin3 is a powerful tool for skin dose calculations when working with beta emitter sources. therefore, it can be utilized in dealing with the issue of radiation protection.
similar resources
Effectiveness Evaluation of Skin Covers against Intravascular Brachytherapy Sources Using VARSKIN3 Code
Background and Objective: The most common intravascular brachytherapy sources include 32P, 188Re, 106Rh and 90Sr/90Y. In this research, skin absorbed dose for different covering materials in dealing with these sources were evaluated and the best covering material for skin protection and reduction of absorbed dose by radiation staff was recognized and recommended.Method: Four materials including...
full textEffectiveness Evaluation of Skin Covers against Intravascular Brachytherapy Sources Using VARSKIN3 Code
BACKGROUND AND OBJECTIVE The most common intravascular brachytherapy sources include (32)P, (188)Re, (106)Rh and (90)Sr/(90)Y. In this research, skin absorbed dose for different covering materials in dealing with these sources were evaluated and the best covering material for skin protection and reduction of absorbed dose by radiation staff was recognized and recommended. METHOD Four material...
full textSkin Dosimetry of Some Beta Sources using the VARSKIN2 Code
Introduction: Skin dose and depth dose distribution of some beta sources that can contaminate skin or protective clothing covering the skin have been calculated using VARSKIN2. Materials and Methods: In this research, absorbed dose to skin and depth dose from the skin surface have been calculated for 90Sr/90Y, 147Pm, 32P beta emitter radioisotopes by using input parameters: activity of source, ...
full textskin dosimetry of some beta sources using the varskin2 code
introduction: skin dose and depth dose distribution of some beta sources that can contaminate skin or protective clothing covering the skin have been calculated using varskin2. materials and methods: in this research, absorbed dose to skin and depth dose from the skin surface have been calculated for 90sr/90y, 147pm, 32p beta emitter radioisotopes by using input parameters: activity of source, ...
full textAccuracy Evaluation of Oncentraâ„¢ TPS in HDR Brachytherapy of Nasopharynx Cancer Using EGSnrc Monte Carlo Code
Background: HDR brachytherapy is one of the commonest methods of nasopharyngeal cancer treatment. In this method, depending on how advanced one tumor is, 2 to 6 Gy dose as intracavitary brachytherapy is prescribed. Due to high dose rate and tumor location, accuracy evaluation of treatment planning system (TPS) is particularly important. Common methods used in TPS dosimetry are based on computat...
full textCalibration and characterization of beta-particle sources for intravascular brachytherapy.
The calibration of a catheter-based system to be used for therapeutic radiation treatment to prevent restenosis following interventional coronary procedures is described. The primary dosimetry was performed ionometrically using an extrapolation chamber equipped with a 1-mm diameter collecting electrode to measure absorbed dose in tissue equivalent plastic at a depth of 2 mm. These results are c...
full textMy Resources
Save resource for easier access later
Journal title:
journal of biomedical physics and engineeringجلد ۳، شماره ۴ Dec، صفحات ۰-۰
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023