quantitative evaluation of lif: mg, ti (tld-100) long-term sensitivity stability

Authors

f mianji

s baradaran

abstract

background : sensitivity is an important dosimetric issue particularly in personal dosimetry where the range of doses is often very low to medium. in-service aging is assumed to be of negligible effect on the response of thermoluminescent dosimeters (tld), consequently, some service providers may mix fresh tlds to in-use ones and simply normalize the readings of tlds through batch calibration. this may lead to an increase in the coefficient of variation of the batch. objective: the correlation between the sensitivity of lif(mg:ti) and its in-service life is quantitatively studied in this paper.  method: three groups of dosimeters with different service lives were chosen from the inventory of the national tld service provider. the tlds were then exposed to different doses and their tl responses were measured and compared.  result: by analyzing the results obtained from groups with different in-service ages, it was shown that the sensitivity of the dosimeters degrades with age in long term, and the amount of this sensitivity loss was measured. the reading standard deviation of the dosimeters degrades with the service life for low doses but seems to be insensitive to the service life in medium and high doses. conclusion: a more efficient annealing approach is recommended time to time for tlds. moreover, the calibration dosimeters must be updated whenever a big difference between the sensitivity of new and used tlds is seen.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

Investigation of LiF, Mg and Ti (TLD-100) Reproducibility

LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the values of Element Correction Coefficient (ECC) were obtained 4 times by irradiating them to doses...

full text

Investigation of LiF, Mg and Ti (TLD-100) Reproducibility

LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the values of Element Correction Coefficient (ECC) were obtained 4 times by irradiating them to doses...

full text

investigation of lif, mg and ti (tld-100) reproducibility

lif, mg and ti cubical tld chips (known as tld-100) are widely used for dosimetry purposes. the repeatability of tl dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mgy) with 662kev photons of cs-137. a group of 40 cubical tld chips was randomly selected from a batch and the values of element correction coefficient (ecc) were obtained 4 times by irradiating them to doses...

full text

Evaluation of the Entrance Skin Dose in Animals Undergoing Diagnostic Radiology Using LiF, Mg, Ti Thermoluminescence Dosimetry (TLD-100)

Introduction According to the International Commission on Radiological Protection publication numbers 60 and 21, the environmental control standards should ensure human and other species protection to a desirable degree. Since application of radiographic procedures in Veterinary Medicine has increased significantly, in this study, we aimed to evaluate the entrance skin dose to the animals (e.g....

full text

تولید دزیمتر LiF: Mg, Ti و مقایسه پاسخ آن با دزیمتر (LiF: Mg, Ti (TLD-100 ساخت شرکت هارشا در برابر پرتوهای گاما

Thermoluminescence dosimeters are small tablets used with 3. 2*3. 2*0. 9mm^3 for measurement of received dose from radioactive various beams. The most common dosimeter is TLD-100 which is lithium-fluoride family and this dosimeter contain magnesium impurities. In this study, first lithium-flouride powder was mixed with titanium and controlled atmosphere samples were heated. After measuring ...

full text

evaluation of the entrance skin dose in animals undergoing diagnostic radiology using lif, mg, ti thermoluminescence dosimetry (tld-100)

introduction according to the international commission on radiological protection publication numbers 60 and 21, the environmental control standards should ensure human and other species protection to a desirable degree. since application of radiographic procedures in veterinary medicine has increased significantly, in this study, we aimed to evaluate the entrance skin dose to the animals (e.g....

full text

My Resources

Save resource for easier access later


Journal title:
journal of biomedical physics and engineering

جلد ۲۰۱۲، شماره ۱۲، صفحات ۰-۰

Keywords
[ ' a g i n g ' , ' l i f ( m g ' , ' t i ) ' , ' p e r s o n a l d o s i m e t r y ' , ' s e n s i t i v i t y ' , ' t l d ' , 1 0 0 ]

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023