cu(ii) ion- selective membrane electrode based on bis(n-salicylidene-3,3'-diamino diphenyl) sulfone

Authors

m. aghaie

m. h. fekri

m. giahi

e. baghdar

s. farhadi

abstract

in order to study the cu(ii) ion selective electrode, an optimized membrane was prepared by mixing of 30 mg of pvc powdered, 5 mg of considered ionophore, 60 mg of plasticizer and 5 mg of additive. the sensor showed a near nernstian response for cu(ii) ions over a concentration range from 1.0×0-6 –1.0×0-2 m with a slope of 28.90.5 mv per concentration decade with a working ph range of 3.5-6.3. the limit of detection was 1.010-6 m. the response time of the electrode is about 15s and can be used for more than 2 month without any divergence in potential.    the proposed sensor showed fairly a good discriminating ability towards cu(ii) ion in comparison to other ions. the practical application of the electrode was tested by using it as an indicator electrode to determine the end point in the potentiometric titration of cu(ii) with edta solution.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

Cu(II) Ion- Selective Membrane Electrode Based on bis(N-salicylidene-3,3'-diamino diphenyl) sulfone

In order to study the Cu(II) ion selective electrode, an optimized membrane was prepared by mixing of 30 mg of PVC powdered, 5 mg of considered ionophore, 60 mg of plasticizer and 5 mg of additive. The sensor showed a near Nernstian response for Cu(II) ions over a concentration range from 1.0×0-6 –1.0×0-2 M with a slope of 28.90.5 mV per concentration decade with a working pH range of 3.5-6.3. ...

full text

cu(ii) ion- selective membrane electrode based on bis(n-salicylidene-3,3'-diamino diphenyl) sulfone

in order to study the cu(ii) ion selective electrode, an optimized membrane was prepared by mixing of 30 mg of pvc powdered, 5 mg of considered ionophore, 60 mg of plasticizer and 5 mg of additive. the sensor showed a near nernstian response for cu(ii) ions over a concentration range from 1.0×0-6 –1.0×0-2 m with a slope of 28.90.5 mv per concentration decade with a working ph range of 3.5-6.3. ...

full text

Thermodynamic study of Cu(II) ion-selective membrane electrode based on his (N-salicylidene-3,32-diamino dip hen y sulfone

The effect or temperature on the behavior of Cu(II) Mr electrode based on bis(N-sahcylidenc-3.3'-4iaminodiphenyl) sulfune(BSDDS) was studied. This electrode showed a good Nemstian response in the temperaturerange of 20-50 C. Also, the behavior of the synthesized electrode was inngigated in the mixed solvents'ethanol+water" and "dioxane +water'.

full text

Steroid acid amides of diamino diphenyl sulfone.

For the past several years, I synthesized various new steroid compounds and steroid derivatives for chemotherapeutic purposes. The series of the steroid acid amides of diamino diphenyl sulfone included 4-cholylamino4’-amino diphenyl sulfone (B 110) and 4-desoxycholylamino 4’-amino diphenyl sulfone (Bill) which were chosen among representative members cf this group for more detailed study. A bri...

full text

Selective Membrane Electrode for Bromide Ion Based on Aza Pyrilium Ion Derivative as a new Ionophore.

A highly selective electrode for Bromide ion based on aza pyrilium derivative as an excellentionophore is described.The sensor exhibits a good linear response with a slope of ( 60±1 ) mV per decadeover the concentration range of ( 1×10-3 – 9×10-6 M ) , and a detection limit of ( 3×10-6 M ) of Bromideions .The electrode response is independent of pH in the range of(4.0 –9.5).Selectivity coeffici...

full text

Sn (II) Ion- Selective Membrane Electrode based on Dibenzo-18-Crown-6

We found that dibenzo-18-crown-6 (DB18C6) can be used as an ionophore to make a tin (II)-ionselectivemembrane electrode. The electrode exhibits a near Nernstian response for Sn2+ ions over aconcentration range from 1.0× 10-6 – 1.0× 10-2 M with a slope of 27.5 ± 0.6 mV per concentration decadein an acidic solution (pH=1). The limit of detection was 8.0× 10-7 M. It has a response time of < 20s an...

full text

My Resources

Save resource for easier access later


Journal title:
journal of physical & theoretical chemistry

ISSN

volume 4

issue 1 2007

Hosted on Doprax cloud platform doprax.com