abstract

introduction: well type ionization chamber is a measuring device which is used to determine the activity of brachytherapy sources. the chamber has a cylindrical volume in which a cylindrical tube is mounted in the middle of the chamber. for the measurements, the brachytherapy sources are transferred to the middle of the tube. materials and methods: for designing the well type chamber, the measurement principals of well type chambers were considered and mcnp-4c code as a calculation tool was used. the designed chamber was simulated and the response of the chamber was evaluated. in this investigation, the chamber operational parameters such as operating voltage, leakage current, reproducibility, reference measuring point, recombination and polarization factors as well as response stability for 137cs, 57co and 241am sources were studied. results: the chamber leakage currents at the operating voltage in comparison to the chamber response for the measurement of the above mentioned sources were negligible. the responses of the fabricated chamber for these sources are reproducible and its reference measurement position for these sources was obtained at 6 cm from the bottom of the chamber. the recombination factor for the well type chamber was negligible and the polarization factor is close to 1. therefore, these two factors were not considered in the measurements. the reproducibility of the measurements in different intervals shows the stability of the chamber response for each source. also the results of the chamber current measurement in term of source strength were compared to the response of the simulated chamber for different source positions and energy ranges of the used sources. discussion and conclusion: the results show that the measurement of the reference positions for each source in the simulated and fabricated chamber is quite in a good agreement. regarding the reliable operational properties of the fabricated chamber, this chamber can be used to determine the activities of brachytherapy sources in which the energy ranges are greater than 50 kev.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

Fabrication of Free Air Well Type Ionization Chamber and Calculational Assessment and Measurement of Its Operational Characteristics

Introduction: Well type ionization chamber is a measuring device which is used to determine the activity of brachytherapy sources. The chamber has a cylindrical volume in which a cylindrical tube is mounted in the middle of the chamber. For the measurements, the brachytherapy sources are transferred to the middle of the tube. Materials and Methods: For designing the well type chamber, the measu...

full text

the study of practical and theoretical foundation of credit risk and its coverage

پس از بررسی هر کدام از فاکتورهای نوع صنعت, نوع ضمانت نامه, نرخ بهره , نرخ تورم, ریسک اعتباری کشورها, کارمزد, ریکاوری, gdp, پوشش و وثیقه بر ریسک اعتباری صندوق ضمانت صادرات ایران مشخص گردید که همه فاکتورها به استثنای ریسک اعتباری کشورها و کارمزد بقیه فاکتورها رابطه معناداری با ریسک اعتباری دارند در ضمن نرخ بهره , نرخ تورم, ریکاوری, و نوع صنعت و ریسک کشورها اثر عکس روی ریسک اعتباری داردو پوشش, وثی...

15 صفحه اول

assessment of deep word knowledge in elementary and advanced iranian efl learners: a comparison of selective and productive wat tasks

testing plays a vital role in any language teaching program. it allows teachers and stakeholders, including program administrators, parents, admissions officers and prospective employers to be assured that the learners are progressing according to an accepted standard (douglas, 2010). the problems currently facing language testers have both practical and theoretical implications but the first i...

diagnostic and developmental potentials of dynamic assessment for writing skill

این پایان نامه بدنبال بررسی کاربرد ارزیابی مستمر در یک محیط یادگیری زبان دوم از طریق طرح چهار سوال تحقیق زیر بود: (1) درک توانایی های فراگیران زمانیکه که از طریق برآورد عملکرد مستقل آنها امکان پذیر نباشد اما در طول جلسات ارزیابی مستمر مشخص شوند; (2) امکان تقویت توانایی های فراگیران از طریق ارزیابی مستمر; (3) سودمندی ارزیابی مستمر در هدایت آموزش فردی به سمتی که به منطقه ی تقریبی رشد افراد حساس ا...

15 صفحه اول

My Resources

Save resource for easier access later


Journal title:
iranian journal of medical physics

جلد ۴، شماره Issue ۳,۴، صفحات ۱-۱۰

Hosted on Doprax cloud platform doprax.com

copyright © 2015-2023