an assessment of spinal cord dose following radiotherapy of nasopharyngeal cancer by tld and rando phantom

Authors

mohammad taghi bahreyni toossi professor, medical physics research center, faculty of medicine, mashhad university of medical sciences, mashhad, iran

eftekhar rajab bolokat m.sc., in medical physics, mashhad university of medical sciences, mashhad, iran

raham salek m.sc., in medical engineering, medical physics research center, faculty of medicine, mashhad university of medical sciences, mashhad, iran

shahram baiani associate professor, radiation oncology dept., omid hospital, mashhad university of medical sciences, mashhad, iran

abstract

introduction: nasopharyngeal carcinoma is one of the most common malignancies in the head and neck region and radiotherapy is its treatment of choice. in spite of the fact that it is widely used, due to the presence of many sensitive organs or tissues in this region, patients may suffer from a wide range of side effects. one such sensitive tissue is the spinal cord. if the absorbed dose to spinal cord is greater than its tolerance dose, then myelopathy and lhermitte’s sign are not avoidable. material and methods: thehead and neck of a rando phantom (reference man) was employed as a hypothetical patient suffering from nasopharyngeal carcinoma. the full course of treatment consisted of three phases. at the beginning of every phase, an oncologist used a simulator to delineate the surface of the rando phantom for treatment. tld chips (tld-100) were employed for dose measurement. tld chips were inserted in the previously made holes on the surface of selected slices adjacent to second cervical to fourth thoracic vertebra. absorbed dose by tlds were read by a harshaw 3500 tld reader. results: total measured dose (in gy) of various parts of spinal cord adjacent to second cervical to fourth thoracic vertebra varied widely and were as follows respectively: 15.24±1.31, 50.31±1.06, 49.15±2.77, 47.48±1.42, 54.56±2.6, 48.92±0.6, 45.1±0.45. in other words, the range of doses received by different segments of the spinal cord could be as wide as 15.24 to 54.56 gy. conclusion: although the spinal cord was excluded at the end of the first phase, a significant change in the absorbed dose at the end of the first and second phases was not observed. in phase three, the anterior neck field was replaced by a lateral field and the spinal cord absorbed dose was reduced considerably. according to our results, absorbed doses of the spinal cord segments corresponding to the region confined between the third cervical to third thoracic vertebra were more than the 47 gy recommended tolerance dose value. therefore, special attention must be paid to protect this sensitive tissue while the treatment is performed. application of modern techniques such as imrt, if available, will reduce the unnecessary dose the spinal cord and its consequent biological risks considerably.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

An Assessment of Spinal Cord Dose Following Radiotherapy of Nasopharyngeal Cancer by TLD and Rando Phantom

Introduction: Nasopharyngeal carcinoma is one of the most common malignancies in the head and neck region and radiotherapy is its treatment of choice. In spite of the fact that it is widely used, due to the presence of many sensitive organs or tissues in this region, patients may suffer from a wide range of side effects. One such sensitive tissue is the spinal cord. If the absorbed dose to spin...

full text

Dose Measurements of Parotid Glands and Spinal Cord in Conventional Treatment of Nasopharyngeal Carcinoma Using RANDO Phantom and Thermoluminescent Dosimeters

Introduction Radiotherapy is regarded as the first treatment of choice for nasopharyngeal carcinoma. Despite the advantages of radiotherapy, patients may suffer from a wide range of side-effects due to the presence of many sensitive normal tissues in these regions. If the absorbed dose exceeds the tolerance level in parotid glands and the spinal cord, myelopathy, Lhermitte's sign and xerostomia...

full text

Dose Measurements of Parotid Glands and Spinal Cord in Conventional Treatment of Nasopharyngeal Carcinoma Using RANDO Phantom and Thermoluminescent Dosimeters

Introduction Radiotherapy is regarded as the first treatment of choice for nasopharyngeal carcinoma. Despite the advantages of radiotherapy, patients may suffer from a wide range of side-effects due to the presence of many sensitive normal tissues in these regions. If the absorbed dose exceeds the tolerance level in parotid glands and the spinal cord, myelopathy, Lhermitte's sign and xerostomia...

full text

dose measurements of parotid glands and spinal cord in conventional treatment of nasopharyngeal carcinoma using rando phantom and thermoluminescent dosimeters

introduction radiotherapy is regarded as the first treatment of choice for nasopharyngeal carcinoma. despite the advantages of radiotherapy, patients may suffer from a wide range of side-effects due to the presence of many sensitive normal tissues in these regions. if the absorbed dose exceeds the tolerance level in parotid glands and the spinal cord, myelopathy, lhermitte's sign and xeros...

full text

Measurement of Thyroid Dose by TLD arising from Radiotherapy of Breast Cancer Patients from Supraclavicular Field

Background: Breast cancer is the most frequently diagnosed cancer and the leading global cause of cancer death among women worldwide. Radiotherapy plays a significant role in treatment of breast cancer and reduces locoregional recurrence and eventually improves survival. The treatment fields applied for breast cancer treatment include: tangential, axillary, supraclavicular and internal mammary ...

full text

Evaluation of Absorbed Dose of Critical Organ in Rando Phantom under Head, Abdomen and Pelvis Spiral CT Scan by Thermo Luminescent Dosimetery - TLD

Background & Objectives: Computed tomography (CT) represents 11% of all diagnostic radiology procedures but it contributes to almost 67% of the total effective dose to the human population. In head and neck CT which consist of 1/3 of total CT scans, other critical organs such as lenses and thyroid are in the radiation field. Also in the abdomen and pelvis scan, irradiation of ovaries is unavoid...