an investigation on the performance of dose calibrators in nuclear medicine centers in iran
Authors
abstract
background: to investigate the status of the nuclear medicine (nm) centers in iran for the performance of dose calibrators, 18 out of 54 centers providing nm services in iran were randomly selected and inspected in 1997. in the first phase of the study the selected centers were inspected for performing of quality control (qc) tests of dose calibrators. the linearity of the activity response, precision, accuracy, and the physical functions of the instruments, were studied. in the second phase of the study, carried out in 2006, 28 out of 75 nm centers were investigated for qc tests performance. materials and methods: the qc tests were performed by using standardized radio nuclides of tc-99m and cs-137 in the first phase, and tc-99m and i-131 in the second phase of the studies. standard procedures were used for carrying out the tests. results: according to the obtained results in the first phase of the study, 10 centers were found to be in unacceptable situation. following this study, all the concerned nm centers were informed about the results, and at the same time the repair and adjustment of the dose calibrators were requested. in addition, the appropriate training courses along with the qc testing manuals were provided to the centers. based on the data of the second phase of the study, only 6 nm centers were in unacceptable situation. the results indicated the effectiveness of the improvements carried out in the working procedures of the centers during interval between the two phases of investigation.
similar resources
An investigation on the performance of dose calibrators in nuclear medicine centers in Iran
Background: To investigate the status of the nuclear medicine (NM) centers in Iran for the performance of dose calibrators, 18 out of 54 centers providing NM services in Iran were randomly selected and inspected in 1997. In the first phase of the study the selected centers were inspected for performing of quality control (QC) tests of dose calibrators. The linearity of the activity res...
full textthe impact of e-readiness on ec success in public sector in iran the impact of e-readiness on ec success in public sector in iran
acknowledge the importance of e-commerce to their countries and to survival of their businesses and in creating and encouraging an atmosphere for the wide adoption and success of e-commerce in the long term. the investment for implementing e-commerce in the public sector is one of the areas which is focused in government‘s action plan for cross-disciplinary it development and e-readiness in go...
an investigation on influencing factors on tourists shopping attitude of iranian handmade carpet in isfahan
چکیده ندارد.
15 صفحه اولQuality Control of Radiopharmaceutical Dose Calibrators in Nuclear Medicine Unit
As part of the program to ensure quality in nuclear medicine unit, in addition to diagnostic procedures, are evaluated activity meters, which is intended to measure the aliquot of radiation of radionuclides and / or radiopharmaceuticals that are administered to patients undergoing diagnostic investigation and / or therapeutic treatment. The good operating condition of dose calibrators is essent...
full textan investigation about the relationship between insurance lines and economic growth; the case study of iran
مطالعات قبلی بازار بیمه را به صورت کلی در نظر می گرفتند اما در این مطالعه صنعت بیمه به عنوان متغیر مستفل به بیمه های زندگی و غیر زندگی شکسته شده و هم چنین بیمه های زندگی به رشته های مختلف بیمه ای که در بازار بیمه ایران سهم قابل توجهی دارند تقسیم میشود. با استفاده از روشهای اقتصاد سنجی داده های برای دوره های 48-89 از مراکز ملی داده جمع آوری شد سپس با تخمین مدل خود بازگشتی برداری همراه با تعدادی ...
15 صفحه اولMy Resources
Save resource for easier access later
Journal title:
iranian journal of radiation researchجلد ۶، شماره ۲، صفحات ۶۴-۶۹
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023