verifying the accuracy of dose distribution in gamma knife unit in presence of inhomogeneities using pagat polymer gel dosimeter and mc simulation

Authors

t. allahverdi pourfallah department of medical physics and biomedical engineering, tehran university of medical sciences, tehran, iran

m. allahverdi department of medical physics and biomedical engineering, tehran university of medical sciences, tehran, iran

n. riyahi alam department of medical physics and biomedical engineering, tehran university of medical sciences, tehran, iran

m.r. ay research center for science and technology in medicine, tehran university of medical sciences, tehran, iran

abstract

background: polymer gel dosimetry is still the only dosimetry method for direct measuring of threedimensional dose distributions. mri polymer gel dosimeters are tissue equivalent and can act as a phantom material. in this study the obtained isodose maps with pagat polymer gel dosimeter were compared to those calculated with egsnrs for singleshot irradiations of 8 and 18 mm collimators of gamma knife (gk) unit in homogeneous and inhomogeneous phantoms. materials and methods: a custom-built, 16 cm diameter spherical plexiglas head phantom was. inside the phantom, there was one cubic cutout for insertion of gel phantoms, and another cutout for inserting the inhomogeneities. the phantoms were scanned with a siemens clinical 1.5 t mri scanner. the multiple spin-echo sequence with 32 echoes was used for the mri scans. results: the results of measurement and simulation in homogeneous and inhomogeneous phantoms showed that the presence of inhomogeneities in head phantom could cause spatial uncertainty higher than ±2 mm and dose uncertainty higher than 7%. conclusion: the presence of inhomogeneities could cause dose differences which were not in accordance with accuracy in treatment with gk radiosurgery. moreover, the findings of monte carlo calculation revealed that the applied simulation code (egsnrc) was a proper tool for evaluation of 3d dose distribution in gk unit. iran. j. radiat. res., 2009 7 (1): 49-56

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

Verifying the accuracy of dose distribution in Gamma Knife unit in presence of inhomogeneities using PAGAT polymer gel dosimeter and MC simulation

Background: Polymer gel dosimetry is still the only dosimetry method for direct measuring of threedimensional dose distributions. MRI Polymer gel dosimeters are tissue equivalent and can act as a phantom material. In this study the obtained isodose maps with PAGAT polymer gel dosimeter were compared to those calculated with EGSnrs for singleshot irradiations of 8 and 18 mm collimators of...

full text

Verifying the accuracy of dose distribution in Gamma Knife unit in presence of inhomogenei- ties using PAGAT polymer gel dosimeter and MC simulation

*Corresponding author: Dr. Tayyeb Allahverdi Pourfallah, Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran. Fax: +98 21 88973653 E-mail: [email protected] Background: Polymer gel dosimetry is still the only dosimetry method for direct measuring of threedimensional dose distributions. MRI Polymer gel dosimeters are tissue equi...

full text

Evaluation of the Effects of Inhomogeneities on Dose Profiles Using Polymer Gel Dosimeter and Monte Carlo Simulation in Gamma Knife

Introduction Polymer gel dosimeters offer a practical solution to 3D dose verification for conventional radiotherapy as well as intensity-modulated and stereotactic radiotherapy. In this study, EGSnrc calculated and PAGAT polymer gel dosimeter measured dose profiles from single shot irradiation with 18 mm collimator of Gamma Knife in homogeneous and inhomogeneous phantoms were compared with eac...

full text

evaluation of the effects of inhomogeneities on dose profiles using polymer gel dosimeter and monte carlo simulation in gamma knife

introduction polymer gel dosimeters offer a practical solution to 3d dose verification for conventional radiotherapy as well as intensity-modulated and stereotactic radiotherapy. in this study, egsnrc calculated and pagat polymer gel dosimeter measured dose profiles from single shot irradiation with 18 mm collimator of gamma knife in homogeneous and inhomogeneous phantoms were compared with eac...

full text

Evaluation of the Effects of Inhomogeneities on Dose Profiles Using Polymer Gel Dosimeter and Monte Carlo Simulation in Gamma Knife

Introduction Polymer gel dosimeters offer a practical solution to 3D dose verification for conventional radiotherapy as well as intensity-modulated and stereotactic radiotherapy. In this study, EGSnrc calculated and PAGAT polymer gel dosimeter measured dose profiles from single shot irradiation with 18 mm collimator of Gamma Knife in homogeneous and inhomogeneous phantoms were compared with eac...

full text

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Background: In this work investigation of the normoxic PAGAT polymer gel dosimeter such as sensitivity, the R2-dose response with post time and the percentage depth dose (PDD) of PAGAT polymer gel dosimeter have been undertaken. Materials and Methods: Using MRI, the formulation to give the maximum change in the transverse relaxation rate R2 was determined to be 4.5% N,N'-methylenbis- ...