calibration of iridium-192 source by ionization chamber for high dose rate brachytherapy
Authors
abstract
background: the effectiveness and safety of brachytherapy treatment is mainly concerned with the calibration of sources and their traceability to internationally accepted standards. secondary standard dosimetry laboratory (ssdl) does not offer calibration of ionization chambers with gamma-ray spectrum of high dose rate source 192ir . this work has been carried out to calibrate the high dose rate (hdr) 192ir sources. materials and methods: an interpolation procedure, using calibrations above 1.25 mev (60co), and below 135 kv x-rays (61.1 kev), the exposure-weighted average energy 397 kev of 192ir is employed. using farmer ionization chamber hdr 192ir source has been calibrated by free in-air measurement technique and then the calibrated source has been used to calibrate well-type ionization chamber. results: difference between in-air measurement and that of manufacturer’s remained within ± 3%, and the difference between in-air measurement and well-type chamber remained within ± 1%. comparison between well-type measurements and manufacturer values shows differences less than ± 2%. all these differences are within the acceptable tolerance limits. conclusion: the method presented shows good degree of accuracy so that the traceability of calibration is assured. installation of a new brachytherapy source needs to be calibrated by in-air measurement technique and then this calibrated source should be used for well chamber calibration. brachytherapy community is suggested to calibrate their sources with this method.
similar resources
Calibration of iridium-192 source by ionization chamber for high dose rate brachytherapy
Background: The effectiveness and safety of brachytherapy treatment is mainly concerned with the calibration of sources and their traceability to internationally accepted Standards. Secondary Standard Dosimetry Laboratory (SSDL) does not offer calibration of ionization chambers with gamma-ray spectrum of high dose rate source 192Ir . This work has been carried out to calibrate the high dose rat...
full textDosimetric characterization of the M−15 high‐dose‐rate Iridium−192 brachytherapy source using the AAPM and ESTRO formalism
The Source Production & Equipment Co. (SPEC) model M-15 is a new Iridium-192 brachytherapy source model intended for use as a temporary high-dose-rate (HDR) brachytherapy source for the Nucletron microSelectron Classic afterloading system. The purpose of this study is to characterize this HDR source for clinical application by obtaining a complete set of Monte Carlo calculated dosimetric pa...
full textInverse Planning in Brachytherapy: Radium to High Dose Rate 192 Iridium Afterloading
We consider the inverse planning problem in brachytherapy, i.e. the problem to determine an optimal number of catheters, number of sources for low-dose rate brachytherapy (LDR) and the optimal dwell times for high-dose rate brachytherapy (HDR) necessary to obtain an optimal as possible dose distribution. Starting from the 1930s, inverse planning for LDR brachytherapy used geometrically derived ...
full textIndependent assessment of source transit time for the BEBIG SagiNova® high dose rate brachytherapy afterloader
Introduction: The dwell time and transit time components contribute to the overall delivered dose to patients in high dose rate (HDR) brachytherapy treatments. The transit time results from source entry and exit as well as source movements between dwell positions. It depends on various parameters such as the source speed profile, source indexer distance to dwell position, and ...
full textDosimetric investigation of high dose rate Ir-192 source with Monte Carlo method
Background: This study aims to calculate the air-kerma strength (SK), the dose rate constant (∧) and the dose rate profiles of a Gammamed 12i Ir-192 source by using the Monte Carlo technique and to compare the dose rate values with those calculated by the Abacus HDR treatment planning system (TPS). Materials and Methods: Air-kerma strength (in units of U; 1 U= μGy m2 h-1) and the dose ra...
full textSingle dose irradiation response of pig skin: a comparison of brachytherapy using a single, high dose rate iridium-192 stepping source with 200 kV X-rays.
An experimental brachytherapy model has been developed to study acute and late normal tissue reactions as a tool to examine the effects of clinically relevant multifractionation schedules. Pig skin was used as a model since its morphology, structure, cell kinetics and radiation-induced responses are similar to human skin. Brachytherapy was performed using a microSelectron high dose rate (HDR) a...
full textMy Resources
Save resource for easier access later
Journal title:
iranian journal of radiation researchجلد ۱۱، شماره ۳، صفحات ۱۸۹-۱۹۳
Keywords
Hosted on Doprax cloud platform doprax.com
copyright © 2015-2023