assessing the effect of dental implants thread design on distribution of stress in impact loadings using three dimensional finite element method

Authors

iman zarei department of mechanical engineering, collage of engineering, shiraz branch, islamic azad university , shiraz, iran

salar khajehpour department of mechanical engineering, collage of engineering, shiraz branch, islamic azad university , shiraz, iran

abbas sabouri department of prosthodontics, school of dentistry, shiraz university of medical sciences, shiraz, iran.

abdol aziz haghnegahdar department of oral and maxillofacial radiology, school of dentistry, shiraz university of medical sciences, shiraz, iran

abstract

statement of problem: impacts and accidents are considered as the main fac- tors in losing the teeth, so the analysis and design of the implants that they can be more resistant against impacts is very important. one of the important nu- merical methods having widespread application in various fields of engineering sciences is the finite element method. among its wide applications, the study of distribution of power in complex structures can be noted. objectives: the aim of this research was to assess the geometric effect and the type of implant thread on its performance; we also made an attempt to determine the created stress using finite element method. materials and methods: in this study, the three dimensional model of bone by using cone beam computerized tomography (cbct) of the patient has been provided. the implants in this study are designed by solid works software. loading is simulated in explicit dynamic, by struck of a rigid body with the speed of 1 mm/s to implant vertically and horizontally; and the maximum level of induced stress for the cortical and trabecular bone in the ansys workbench software was calculated. results: by considering the results of this study, it was identified that, among the designed samples, the maximum imposed stress in the cortical bone layer occurred in the first group (straight threads) and the maximum stress value in the trabecular bone layer and implant occurred in the second group (tapered threads). conclusions: due to the limitations of this study, the implants with more depth thread, because of the increased contact surface of the implant with the bone, caused more stability;   also, the implant with smaller thread and shorter pitch length caused more stress to the bone.

Upgrade to premium to download articles

Sign up to access the full text

Already have an account?login

similar resources

Assessing the Effect of Dental Implants Thread Design on Distribution of Stress in Impact Loadings Using Three Dimensional Finite Element Method

STATEMENT OF PROBLEM Impacts and accidents are considered as the main fac- tors in losing the teeth, so the analysis and design of the implants that they can be more resistant against impacts is very important. One of the important nu- merical methods having widespread application in various fields of engineering sciences is the finite element method. Among its wide applications, the study of d...

full text

Evaluating the impact of length and thread pitch on the stress distribution in dental implants and surrounding bone using finite element method

 longevity of osseointegrated implants are intensely influenced by biomechanical factors. Control of these factors prevents mechanical complications, which include fracture of screws, components, or materials veneering the framework. In this study, the impact of length and threads pitch of dental implants on the stress distribution and maximum Von Mises stress in implant-abutment complex and ja...

full text

evaluating the impact of length and thread pitch on the stress distribution in dental implants and surrounding bone using finite element method

longevity of osseointegrated implants are intensely influenced by biomechanical factors. control of these factors prevents mechanical complications, which include fracture of screws, components, or materials veneering the framework. in this study, the impact of length and threads pitch of dental implants on the stress distribution and maximum von mises stress in implant-abutment complex and jaw...

full text

buckling of viscoelastic composite plates using the finite strip method

در سال های اخیر، تقاضای استفاده از تئوری خطی ویسکوالاستیسیته بیشتر شده است. با افزایش استفاده از کامپوزیت های پیشرفته در صنایع هوایی و همچنین استفاده روزافزون از مواد پلیمری، اهمیت روش های دقیق طراحی و تحلیل چنین ساختارهایی بیشتر شده است. این مواد جدید از خودشان رفتارهای مکانیکی ارائه می دهند که با تئوری های الاستیسیته و ویسکوزیته، نمی توان آن ها را توصیف کرد. این مواد، خواص ویسکوالاستیک دارند....

the impact of using inspirational quotes on abstract vocabulary recall

the present study is an attempt to investigate the potential impact of inspirational quotes on improving english abstract vocabulary recall. to achieve this goal, a multiple choice language proficiency test of 60 items including vocabulary and grammar component was administered to a sample of 63 second-semester male and female students whose age ranged between 17 to 22 and they were studying en...

15 صفحه اول

Influence of three different implant thread designs on stress distribution: A three-dimensional finite element analysis

PURPOSE Clinical success of implant prosthodontics is dependent in part upon the type of implant thread design. The selection of implant thread design plays an important role in the outcome of the treatment. This study was undertaken to evaluate the pattern of stress distribution using a finite element analysis; hence, the area which would be bearing maximum load for a given design would be arr...

full text

My Resources

Save resource for easier access later


Journal title:
journal of dental biomaterials

جلد ۳، شماره ۲، صفحات ۲۳۳-۰