high-performance liquid chromatographic determination of theophylline in human serum
Authors
abstract
a simple, rapid and sensitive hplc method for the determination of theophylline (t) in human serum has been developed. an isocratic system consisting of a /-l bondapak c18 column, mobile phase of methanol, phosphate buffer (22:78, ph=4.5), and a flow rate of 1.4 ml/min was used. the eluent was detected by uv at 275 nm at room temperature. 8-chlorotheophylline (8-ct) was used as an internal standard. the serum samples deproteinated by methanol containing 8-ct and the supernatant was injected into the hplc system. the retention times of 5.7 and 8.1 min were found for t and 8-ct, respectively. the linearity was checked in the range of 0.2- 30 µg/ml. relative standard deviation for both inter-day and intra-day precision analysis was less than 5%. no interference was observed from endogenous serum components. specificity was shown against some commonly co-administered drugs. simple and fast sample preparation, small sample volume (250 µl), precision, reproducibility, specificity, sensitivity and high percentage recovery (98 %) make the method to be practically useful for t monitoring in asthmatic patients.
similar resources
HIGH-PERFORMANCE LIQUID CHROMATOGRAPHIC DETERMINATION OF THEOPHYLLINE IN HUMAN SERUM
A simple, rapid and sensitive HPLC method for the determination of theophylline (T) in human serum has been developed. An isocratic system consisting of a /-l Bondapak C18 column, mobile phase of methanol, phosphate buffer (22:78, pH=4.5), and a flow rate of 1.4 mL/min was used. The eluent was detected by UV at 275 nm at room temperature. 8-Chlorotheophylline (8-CT) was used as an internal ...
full textHigh-performance liquid-chromatographic determination theophylline in plasma.
A high-performance liquid-chromatographic method is reported for monitoring theophylline in plasma. Samples are deproteinized with 2.5 volumes of acetonitrile and the supernates chromatographed on a reversed-phase column. Absorption at 275 nm is monitored. One can accurately measure 1.5 mg of theophylline per liter, in as little as 10 microliter of plasma, and only about 7 min is required per s...
full textHigh-Performance Liquid Chromatographic Determination of Rivastig-mine in Human Plasma for Application in Pharmacokinetic Studies
A simple and reproducible HPLC method with spectrophotometric detection was developed for determination of rivastigmine in human plasma. Liquid-liquid extraction of rivastigmine and donepezil (as internal standard) from plasma samples was performed with 1-butanol/n-hexane (2:98 v/v) in alkaline condition followed by back-extraction into diluted acetic acid. Chromatography was carried out using ...
full textHigh Performance Liquid Chromatographic Determination of Inactive Carboxylic Acid Metabolite of Clopidogrel in Human Serum: Application to a Bioequivalence Study
A sensitive and rapid method is described for determination of clopidogrel carboxylic acid (CCA), the inactive metabolite of the antiplatelet agent clopidogrel in human serum. The analytical procedure involves liquid-liquid extraction of the analyte and an internal standard (phenytoin) with ethyl acetate. A mobile phase consisting of 0.05 M phosphate buffer containing triethylamine (0.5...
full textHigh-Performance Liquid Chromatographic Determination of Rivastig-mine in Human Plasma for Application in Pharmacokinetic Studies
A simple and reproducible HPLC method with spectrophotometric detection was developed for determination of rivastigmine in human plasma. Liquid-liquid extraction of rivastigmine and donepezil (as internal standard) from plasma samples was performed with 1-butanol/n-hexane (2:98 v/v) in alkaline condition followed by back-extraction into diluted acetic acid. Chromatography was carried out using ...
full textRapid High Performance Liquid Chromatographic Determination of Risperidone in Human Plasma
A simple, rapid and sensitive high-performance liquid chromatographic (HPLC) method for the determination of risperidone in human plasma was developed. An HPLC system based on a Nucleosil C8 column (150?4 mm) and a UV detector (?= 280 nm) were used. A mixture of sodium dihydrogen phosphate buffer-acetonitrile (55:45, v/v) adjusted to pH 6.0 at a flow rate of 1.5 ml min-1 was used as mobile phas...