Prospective Validation of Intra- and Interobserver Reproducibility of a New Point Shear Wave Elastographic Technique for Assessing Liver Stiffness in Patients with Chronic Liver Disease
نویسندگان
چکیده
OBJECTIVE To assess intra- and inter-observer reproducibility of a new point shear wave elastography technique (pSWE, S-Shearwave, Samsung Medison) and compare its accuracy in assessing liver stiffness (LS) with an established pSWE technique (Virtual Touch Quantification, VTQ). MATERIALS AND METHODS Thirty-three patients were enrolled in this Institutional Review Board-approved prospective study. LS values were measured by VTQ on an Acuson S2000 system (Siemens Healthineer) and S-Shearwave on an RS-80A (Samsung Medison) in the same session, followed by two further S-Shearwave sessions for inter- and intra-observer variation at 8-hour intervals. The technical success rate (SR) and reliability of the measurements of both pSWE techniques were compared. The intra- and inter-observer reproducibility of S-Shearwave was determined by intraclass correlation coefficients (ICCs). LS values were measured by both methods of pSWE. The diagnostic performance in severe fibrosis (F ≥ 3) and cirrhosis (F = 4) was evaluated using the receiver operating characteristics curve analysis and the Obuchowski measure with the LS values of transient elastography as the referenced standard. RESULTS The VTQ (100%, 33/33) and S-Shearwave (96.9%, 32/33) techniques did not display a significant difference in technical SR (p = 0.63) or reliability of LS measurements (96.9%, 32/33; 93.9%, 30/32, respectively, p = 0.61). The inter- and intra-observer agreement for LS measurements using the S-Shearwave technique was excellent (ICC = 0.98 and 0.99, respectively). The mean LS values of both pSWE techniques were not significantly different and exhibited a good correlation (r = 0.78). To detect F ≥ 3 and F = 4, VTQ and S-Shearwave showed comparable diagnostic accuracy as indicated by the following outcomes: areas under receiver operating characteristics curve (AUROC) = 0.87 (95% confidence intervals [CI] 0.70-0.96), 0.89 for VTQ (95% CI 0.74-0.97), respectively; and AUROC = 0.84 (95% CI 0.67-0.94), 0.94 (95% CI 0.80-0.99) for S-Shearwave (p > 0.48), respectively. The Obuchowski measures were similarly high for S-Shearwave and VTQ (0.94 vs. 0.95). CONCLUSION S-Shearwave shows excellent inter- and intra-observer agreement and diagnostic effectiveness comparable to VTQ in detecting LS.
منابع مشابه
Liver fibrosis staging with a new 2D-shear wave elastography using comb-push technique: Applicability, reproducibility, and diagnostic performance
OBJECTIVE To evaluate the applicability, reproducibility, and diagnostic performance of a new 2D-shear wave elastography (SWE) using the comb-push technique (2D CP-SWE) for detection of hepatic fibrosis, using histopathology as the reference standard. MATERIALS AND METHODS This prospective study was approved by the institutional review board, and informed consent was obtained from all patient...
متن کاملPoint shear wave elastography method for assessing liver stiffness.
AIM To estimate the validity of the point shear-wave elastography method by evaluating its reproducibility and accuracy for assessing liver stiffness. METHODS This was a single-center, cross-sectional study. Consecutive patients with chronic viral hepatitis scheduled for liver biopsy (LB) (Group 1) and healthy volunteers (Group 2) were studied. In each subject 10 consecutive point shear-wave ...
متن کامل2-dimensional shear wave elastography: Interobserver agreement and factors related to interobserver discrepancy
PURPOSE To evaluate the interobserver reproducibility of two-dimensional shear wave elastography (2D-SWE) in measuring liver stiffness (LS) and to investigate factors related to liver 2D-SWE. MATERIALS AND METHODS A prospective study was conducted between August 2011 and August 2012 in rheumatoid arthritis patients who had been treated with methotrexate. Interobserver reproducibility of 2D-SW...
متن کاملComparative study between two point Shear Wave Elastographic techniques: Acoustic Radiation Force Impulse (ARFI) elastography and ElastPQ.
AIM To compare the feasibility and the liver stiffness (LS) values obtained by means of the two available point Shear Wave Elastographic (pSWE) techniques: ARFI elastography and ElastPQ technique. MATERIAL AND METHODS Our study included 176 subjects with or without chronic liver disease, in which LS was evaluated in the same session by means of ARFI elastography and ElastPQ. RESULTS Ten val...
متن کاملReproducibility of shear wave elastography (SWE) in patients with chronic liver disease
The presence of significant fibrosis is an indicator for liver disease staging and prognosis. The aim of the study was to determine reproducibility of real-time shear wave elastography using a hepatic biopsy as the reference standard to identify patients with chronic liver disease. Forty patients with chronic liver disease and 12 normal subjects received shear wave elastography performed by ski...
متن کامل