Regarding Article 'Multiple Episodes of Presyncope in a Pacemaker Dependent Patient: What is the Diagnosis?'
نویسنده
چکیده
NCAP algorithm is designed to prevent competitive atrial pacing in Medtronic dual-chamber pacemakers and is nominally 'on'. It is activated on sensing atrial activity (AR) in the postventricular atrial refractory period. AR event does not trigger a ventricular pace. Also, NCAP algorithm delays the A pace for 300 ms (non-programmable in pacemakers) after AR even though, lower rate or sensor indicated rate may have required atrial pace to occur earlier. This prevents atrial pacing when the atrial tissue is refractory and thus may reduce atrial arrhythmias. The interval 300 ms is chosen based on the observations that an atrial stimulus delivered earlier than 300 ms may find the atrial myocardium refractory. But the present case and another report [2] suggest that the atrial myocardium especially if diseased may not recover excitability in this period. In the NCAP algorithm, the pacemaker also attempts to maintain the ventricular pacing rate at the sensor indicated rate (or the lower rate). This is done by shortening the PAV as is seen in the present example.
منابع مشابه
In Reply 'Multiple Episodes of Presyncope in a Pacemaker Dependent Patient: What is the Diagnosis?'
We thank Dr Bhargava for careful reading of our article [1]. As highlighted by Dr Bhargava, Non Competitive Atrial Pacing (NCAP) algorithm is designed to prevent atrial arrhythmias by preventing atrial pacing within 300msec after a sensed atrial event in the post ventricular atrial refractory period (PVARP). The varying AV delays and fixed interval of 300msec between sensed atrial event (AR) in...
متن کاملMultiple Episodes of Presyncope in a Pacemaker Dependent Patient: What is the Diagnosis?
An 85 year old male presents to the emergency room with presyncopal episode associated with nausea and diaphoresis that lasted for 10 minutes. He has past medical history of hypertension and paroxysmal atrial fibrillation on anticoagulation. He also had complete AV block that necessitated the implantation of a dual chamber pacemaker (Medtronic Kappa KDR 701) 6 years prior to current presentatio...
متن کاملبلوک کامل قلبی مادرزادی، معرفی یک مورد و مرور مقالات مشابه
Abstract: Background and Aim: Since electrocardiography (ECG) is essential for the diagnosis of pediatric asymptomatic Complete Heart Block (CHB) and it is not usually done in the preoperative evaluation of asymptomatic children, physicians might confront with asymptomatic cases of CHB at anesthesia induction. Case Report: In this study, we introduced a 31-month-old boy who faced CHB after anes...
متن کاملThe Improvement of the Atrial Flutter Rhythm upon the Removal of the Infected Permanent Pacemaker Lead
Pacemaker infection has multiple risk factors. Its presentation is most often similar to infected endocarditis and the diagnosis is made through studying blood cultures. Transesophageal echocardiography can confirm the diagnosis. The most common microorganisms are staphylococcus speciesis. As a matter of fact, complete pacemaker removal appears to be the only definite treatment. We presented a ...
متن کاملCardiac pacing for severe childhood neurally mediated syncope with reflex anoxic seizures.
OBJECTIVE To determine whether permanent cardiac pacing could prevent syncope and seizures in children with frequent severe neurally mediated syncope, and if so whether dual chamber pacing was superior to single chamber ventricular pacing. METHODS Dual chamber pacemakers were implanted into 12 children (eight male, four female) aged 2-14 years (median 2.8 years) with frequent episodes of refl...
متن کامل