Incidence of ineffective safety margin testing (<10 J) and efficacy of routine subcutaneous array insertion during implantable cardioverter defibrillator implantation☆

نویسندگان

  • Marc-Alexander Ohlow
  • Marcus Roos
  • Bernward Lauer
  • J. Christoph Geller
چکیده

The purpose of this study was to assess (1) the incidence of safety margin testing <10 J (SMT) and (2) the efficacy/safety of routinely adding a subcutaneous array (SQA) (Medtronic 6996SQ) for these patients. Patients with SMT smaller than a 10-J safety margin from maximum output were considered to have very high readings and underwent SQA insertion. These patients were compared with the rest of the patients who had acceptable SMT (≥10 J). A total of 616 patients underwent ICD implantation during the analysis period. Of those, 16 (2.6%) had SMT <10 J. By univariate analysis, younger age, and non-ischemic cardiomyopathy, were all significant predictors of SMT <10 J (p < 0.05). In all 16 cases, other methods to improve SMT prior to array insertion were attempted but failed for all patients: reversing shock polarity (n = 15), removing the superior vena cava coil (n = 14), reprogramming shock waveform (n = 9), and repositioning right ventricular lead (n = 9). Addition of the SQA successfully increased SMT to within safety margin for all patients (32 ± 2 versus 21 ± 3 J; p < 0.001). Follow-up (mean 48.1 ± 21 months) was available for all patients with SQA, only 2 cases with inappropriate shocks due to atrial fibrillation had to be noted. None of the patients experienced complications due to SQA implantation. SMT <10 J occur in about 2.6% of patients undergoing ICD implantation. SQA insertion corrects this problem without procedural/mid-term complications.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Critical analysis of ineffective post implantation implantable cardioverter-defibrillator-testing

AIM To test of the implantable-cardioverter-defibrillator is done at the time of implantation. We investigate if any testing should be performed. METHODS All consecutive patients between January 2006 and December 2008 undergoing implantable cardioverter-defibrillator (ICD) implantation/replacement (a total of 634 patients) were included in the retrospective study. RESULTS Sixteen patients (...

متن کامل

Intraoperative Defibrillation Testing of Subcutaneous Implantable Cardioverter‐Defibrillator Systems—A Simple Issue?

BACKGROUND The results of the recently published randomized SIMPLE trial question the role of routine intraoperative defibrillation testing. However, testing is still recommended during implantation of the entirely subcutaneous implantable cardioverter-defibrillator (S-ICD) system. To address the question of whether defibrillation testing in S-ICD systems is still necessary, we analyzed the dat...

متن کامل

Implantable cardioverter-defibrillator in a patient with dextrocardia situs inversus

Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...

متن کامل

Inductionless or limited shock testing is possible in most patients with implantable cardioverter- defibrillators/cardiac resynchronization therapy defibrillators: results of the multicenter ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduction Evaluation With Implantable Cardioverter-Defibrillator Implantations).

BACKGROUND Implantable cardioverter-defibrillators and cardiac resynchronization therapy defibrillators have relied on multiple ventricular fibrillation (VF) induction/defibrillation tests at implantation to ensure that the device can reliably sense, detect, and convert VF. The ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduc...

متن کامل

Efficacy and Temporal Stability of Reduced Safety Margins for Ventricular Defibrillation

Background—Traditionally, a safety margin of at least 10 J between the maximum output of the pulse generator and the energy needed for ventricular defibrillation has been used because lower safety margins were associated with unacceptably high rates of failed defibrillation and sudden cardiac death. The Low Energy Safety Study (LESS) was a prospective, randomized assessment of the safety margin...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 16  شماره 

صفحات  -

تاریخ انتشار 2016