Malfunction of the mitral valve prosthesis (Cutter-Smeloff). Clinical and hemodynamic observations in three cases.
نویسندگان
چکیده
Three cases of intermittent obstruction of the mitral prosthesis (Cutter-Smeloff) due to ball variance have been reported. The typical findings were marked variation in the arterial pulse and the intensity of the mitral closing sound and intermittent absence of the opening click. The phonocardiography revealed varying intervals from aortic closing sound to the opening click (A2 OC interval). Hemodynamic studies revealed fluctuating aortic and pulmonary artery wedge pressures in spite of normal sinus rhythm, due to impaction of the ball into the prosthetic ring, opening only with severe elevation of the left atrial pressure. At operation, the balls showed yellowish discoloration, but no disruption of sutures nor thrombosis of the prosthesis was found. The incidence of this phenomenon was 9% with the Cutter-Smeloff mitral prosthesis, but no such case was seen among patients with the Starr-Edwards mitral prosthesis.
منابع مشابه
Malfunction of the Mitral Valve Prosthesis (Cutter-Smeloff)
Three cases of intermittent obstruction of the mitral prosthesis (Cutter-Smeloff) due to ball variance have been reported. The typical findings were marked variation in the arterial pulse and the intensity of the mitral closing sound and intermittent absence of the opening click. The phonocardiography revealed varying intervals from aortic closing sound to the opening click (A2 OC interval). He...
متن کاملSound spectroanalytic diagnosis of malfunctioning prosthetic heart valve.
Sound spectroanalysis was carried out for diagnosis of malfunctioning prosthetic heart valves in 50 cases, which consisted of 31 mitral, 14 aortic, 4 mitral+aortic and 1 tricuspid valve replacements with the age ranging from 17 to 57 years. Types of prosthetic valves used in this series were Starr-Edwards silicon ball valves, metal ball valves, disc valves, Kay-Shiley valves, Smeloff-Cutter val...
متن کاملSuccessful Use of Two Thrombolytic Drugs in Prosthetic Mitral and Aortic Valve Thrombosis
Introduction: Prosthetic valve thrombosis is a rare and severe complication of valve replacement, most often encountered with a mechanical prosthesis. The significant morbidity and mortality associated with this condition warrant rapid diagnostic evaluation. Although surgery is the first-line therapy in symptomatic obstructive mechanical valve thrombosis, thrombolytic therapy has been used as a...
متن کاملSHORT TERM CLINICAL OUTCOME OF PERCUTANEOUS BALLOON MITRAL VALVULOPLASTY VERSUS SURGICAL CLOSED MITRAL COMMISSUROTOMY
BaIlon mitral valvuloplasty and closed mitral valve conunissurotomy were performed in 450 and 127 patients, respectively. Also, 254 patients with severe mitral stenosis were enrolled in a prospective randomized trial comparing the two procedures. Immediately after balloon mitral valvuloplasty, mean left atrial pressure was 11.38±3.54 mmHg, transmitral valve gradient was 1.8±2 mmHg, and mit...
متن کاملAN INNOVATIVE SURGIC AL TECHNIQUE FOR CONTINUOUS SUTURING OF PROSTHESIS IN MITR AL VALVE REPL A CEMENT
Mitral valve replacement (MVR) is usually performed in two methods of running and interrupted suturing. In running method, the suturing of the valve is usually with some traction on the annulus and also there is a risk of aortic valve injury. This study was conducted to evaluate a new technique for suturing to avoid these complications and decrease the time of cardiac arrest. One hundred an...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Circulation
دوره 41 3 شماره
صفحات -
تاریخ انتشار 1970