Ultrasmall volume molecular isothermal amplification in microfluidic chip with advanced surface processing
نویسندگان
چکیده
In this paper, we developed a metal micro-fluidic chip with advanced surface processing for ultra-small volume molecular isothermal amplification. This method takes advantages of the nucleic acid amplification with good stability and consistency, high sensitivity about 31 genomic DNA copies and bacteria specific gene identification. Based on the advanced surface processing, the bioreaction assays of nucleic acid amplification was dropped about 392nl in volume. A high numerical aperture confocal optical detection system was advanced to sensitively monitor the DNA amplification with low noise and high power collecting fluorescence near to the optical diffraction limit. A speedy nucleic acid isothermal amplification was performed in the ultra-small volume microfluidic chip, where the time at the inflexions of second derivative to DNA exponential amplified curves was brought forward and the sensitivity was improved about 65 folds to that of in current 25μl Ep-tube amplified reaction, which indicates a promising clinic molecular diagnostics in the droplet amplification.
منابع مشابه
fM to aM nucleic acid amplification for molecular diagnostics in a non-stick-coated metal microfluidic bioreactor
A sensitive DNA isothermal amplification method for the detection of DNA at fM to aM concentrations for pathogen identification was developed using a non-stick-coated metal microfluidic bioreactor. A portable confocal optical detector was utilized to monitor the DNA amplification in micro- to nanoliter reaction assays in real-time, with fluorescence collection near the optical diffraction limit...
متن کاملAn integrated versatile lab-on-a-chip platform for the isolation and nucleic acid-based detection of pathogens
AIM Processing of the samples in molecular diagnostics is complex and labor-intensive. An integrated and automated platform for sample preparation and nucleic acid-based detection can significantly relieve this burden for the users. RESULTS We present a prototype of a versatile and integrated platform for the detection of pathogens in various liquid media. We describe a proof-of-concept for t...
متن کاملMicrofluidic devices for nucleic acid (NA) isolation, isothermal NA amplification, and real-time detection.
Molecular (nucleic acid)-based diagnostics tests have many advantages over immunoassays, particularly with regard to sensitivity and specificity. Most on-site diagnostic tests, however, are immunoassay-based because conventional nucleic acid-based tests (NATs) require extensive sample processing, trained operators, and specialized equipment. To make NATs more convenient, especially for point-of...
متن کاملIntegrated Microfluidic Nucleic Acid Isolation, Isothermal Amplification, and Amplicon Quantification
Microfluidic components and systems for rapid (<60 min), low-cost, convenient, field-deployable sequence-specific nucleic acid-based amplification tests (NAATs) are described. A microfluidic point-of-care (POC) diagnostics test to quantify HIV viral load from blood samples serves as a representative and instructive example to discuss the technical issues and capabilities of "lab on a chip" NAAT...
متن کاملA magnetic bead-based assay for the rapid detection of methicillin-resistant Staphylococcus aureus by using a microfluidic system with integrated loop-mediated isothermal amplification.
This study reports a new diagnostic assay for the rapid detection of methicillin-resistant Staphylococcus aureus (MRSA) by combing nucleic acid extraction and isothermal amplification of target nucleic acids in a magnetic bead-based microfluidic system. By using specific probe-conjugated magnetic beads, the target deoxyribonucleic acid (DNA) of the MRSA can be specifically recognized and hybrid...
متن کامل