A novel approach for superficial intraoperative radiotherapy (IORT) using a 50 kV X‐ray source: a technical and case report
نویسندگان
چکیده
The use of IORT as a treatment modality for patients with close or positive margins has increased over the past decade. For situations where a flat area (up to 6 cm in diameter) has to be treated intraoperatively, new applicators for superficial treatment with a miniature X-ray source (INTRABEAM system) were developed. Here we report our evaluation of the dosimetric characteristics of these new applicators and their first clinical use. Each of these flat and surface applicators consists of a radiation protective metal tube and a flattening filter, which converts the spherical dose distribution of the X-ray source into a flat one. The homogeneity of each dose distribution and depth-dose measurements were evaluated using film dosimetry in a solid water phantom and a soft X-ray ionization chamber in a water tank. The first patient was treated with 5 Gy delivered in 5 mm using a 4 cm FLAT applicator over 21 minutes. The flat applicators show the maximum homogeneity, with a uniformity ratio of 1.02-1.08 in certain depths. In 1 mm depth surface applicators show a uniformity ratio of 1.15-1.28. They also show a higher dose rate and a steeper dose gradient compared to the flat applicators. The results of this investigation demonstrated that the flat and surface applicators have unique dosimetric characteristics that need to be considered during the treatment planning stages. This work also showed that it is possible to perform a superficial localized IORT which provides new application possibilities for use of the INTRABEAM system.
منابع مشابه
Kypho-IORT - a novel approach of intraoperative radiotherapy during kyphoplasty for vertebral metastases
BACKGROUND Instable and painful vertebral metastases in patients with progressive visceral metastases present a common therapeutic dilemma. We developed a novel approach to deliver intraoperative radiotherapy (IORT) during kyphoplasty and report the first treated case. METHODS/RESULTS 60 year old patient with metastasizing breast cancer under chemotherapy presented with a newly diagnosed pain...
متن کاملIntraoperative radiotherapy (IORT) with 50-kv X-ray machines as boost in breast cancer -- more questions than answers.
متن کامل
Intraoperative radiotherapy (IORT) for breast cancer using the Intrabeam system.
INTRODUCTION Intraoperative radiotherapy (IORT) with low-energy X-rays (30-50 KV) is an innovative technique that can be used both for accelerated partial breast irradiation (APBI) and intraoperative boosting in patients affected by breast cancer. Immediately after tumor resection the tumor bed can be treated with low-distance X-rays by a single high dose. Whereas often a geographic miss in cov...
متن کاملMonte Carlo Simulation of Electron Beams produced by LIAC Intraoperative Radiation Therapy Accelerator
Background: One of the main problems of dedicated IORT accelerators is to determine dosimetric characteristics of the electron beams. Monte Carlo simulation of IORT accelerator head and produced beam will be useful to improve the accuracy of beam dosimetry.Materials and Methods: Liac accelerator head was modeled using the BEAMnrcMonte Carlo simulation system. Phase-space files were generated at...
متن کاملIntraoperative radiotherapy during kyphoplasty for vertebral metastases (Kypho-IORT): first clinical results.
AIMS AND BACKGROUND Kyphoplasty is an effective procedure providing structural stability and pain alleviation in vertebral metastases. To prevent early regrowth, patients typically receive postoperative fractionated radiotherapy, which is associated with long treatment duration. Therefore, we established a new approach to deliver intraoperative radiotherapy during kyphoplasty to shorten the tre...
متن کامل