Vascular response and mechanical integrity of the new biodegradable polymer coated sirolimus-eluting PROLIM stent implanted in porcine coronary arteries.
نویسندگان
چکیده
BACKGROUND Although durable polymer coated drug-eluting stents (DES) are standard care in percutaneous coronary interventions, new stent platforms employing biodegradable polymer based drug delivery are increasingly being used in clinical practice. AIM To evaluate the short- (28 days) and medium-term (90 days) vascular effects of the new biodegradable polymer coated sirolimus-eluting stent - the PROLIM stent. METHODS The objectives of the study were evaluated using standard angiographic and histological methods. In addition, the mechanical integrity of tested stents was assessed using Faxitron imaging. A total of 12 PROLIM stents, 11 biodegradable polymer only coated stents (BPCS), and 12 bare metal stents (BMS) were implanted in the coronary arteries of 16 female non-atheroslerotic domestic swine using an overstretch of 1.1:1.0. RESULTS At 28 days, neointimal proliferation was significantly lower in the PROLIM and BMS stents compared to the BPCS stents (p ≤ 0.05). Interestingly, despite thin neointima found at this time in the PROLIM group, there was a further significant decrease in neointimal formation between 28 and 90 days (p = 0.04). Although a statistically bigger neointima was found in BPCS stents at 28 days compared to the PROLIM and BMS stents, there was a 50% decrease in the neointimal area at 90 days follow-up (p = 0.02) which reached the level seen in other groups. The endothelialisation was completed in all tested stents after 28 days. There was a significant increase of fibrin depositions in the PROLIM treated arteries at 28 days which were resorbed nearly completely at 90 days follow-up. At 28 days, the inflammatory response was found to be numerically higher in the BPCS stents (p = NS) compared to other tested groups. On the contrary, at 90 days follow-up when the degradation process of the polymer had been completed, the inflammatory reaction decreased substantially to the level seen in the PROLIM and BMS stents. Faxitron analysis of the stented arteries revealed no major abnormalities except for isolated strut fractures observed in the mid portions of two BMS stents and one BPCS stent. CONCLUSIONS The PROLIM - a biodegradable polymer coated sirolimus-eluting stent - demonstrates very good short-term and medium-term angiographic and histological results. The lack of 'catch-up phenomenon', fast endothelialisation process, and minimal inflammatory reaction may contribute to favourable clinical outcomes using PROLIM stents.
منابع مشابه
Novel biodegradable polymer-coated, paclitaxel-eluting stent inhibits neointimal formation in porcine coronary arteries.
BACKGROUND Biodegradable polymer-coated stents may have positive effects on arterial healing, and reduce the need for prolonged antiplatelet therapy. AIM To assess the vascular effects of the biodegradable polymer proposed as a stent coating, as well as to evaluate inhibition of intimal hyperplasia by Biodegradable Polymer-Coated Paclitaxel-Eluting Stents (BP-PES, LUC-Chopin, Balton) in porci...
متن کاملNovel sirolimus-eluting stent Prolim® with a biodegradable polymer in the all-comers population: one year clinical results with quantitative coronary angiography and optical coherence tomography analysis
BACKGROUND The aim of this study was to assess the safety and the efficacy of the novel sirolimus-eluting Prolim® stent with a biodegradable polymer in the all-comers population. METHODS We prospectively enrolled all patients with stable coronary artery disease or acute coronary syndrome treated with Prolim® stent between January and December 2013 in two interventional cardiology centers in P...
متن کاملComparison of Vascular Responses Following New‐Generation Biodegradable and Durable Polymer‐Based Drug‐Eluting Stent Implantation in an Atherosclerotic Rabbit Iliac Artery Model
BACKGROUND Incomplete endothelialization is the primary substrate of late stent thrombosis; however, recent reports have revealed that abnormal vascular responses are also responsible for the occurrence of late stent failure. The aim of the current study was to assess vascular response following deployment of biodegradable polymer-based Synergy (Boston Scientific) and Nobori (Terumo) drug-eluti...
متن کاملIn vitro and in vivo characterization of novel biodegradable polymers for application as drug-eluting stent coatings.
We have used a series of in vitro and in vivo tests to assess the suitability of two new degradable polymers for application as coatings for drug-eluting stents. The first is a family of urethane-linked multi-block copolymers (MBCP) that comprise blocks of lactide, glycolide, epsilon-caprolactone and/or poly(ethylene glycol) chain-extended with 1,4-butanediisocyanate (SynBiosys polymers). The s...
متن کاملInhibition of neointima formation by a novel drug-eluting stent system that allows for dose-adjustable, multiple, and on-site stent coating.
OBJECTIVE The risk of in-stent restenosis can be considerably reduced by stents eluting cytostatic compounds. We created a novel drug-eluting stent system that includes several new features in the rapidly evolving field of stent-based drug delivery. METHODS AND RESULTS The aim of the present study was the preclinical evaluation of a stent-coating system permitting individual, on-site coating ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Kardiologia polska
دوره 70 7 شماره
صفحات -
تاریخ انتشار 2012