Computer-assisted and conventional total knee replacement: a comparative, prospective, randomised study with radiological and CT evaluation.
نویسندگان
چکیده
After obtaining informed consent, 80 patients were randomised to undergo a navigated or conventional total knee replacement. All received a cemented, unconstrained, cruciate-retaining implant with a rotating platform. Full-length standing and lateral radiographs and CT scans of the hip, knee and ankle joint were carried out five to seven days after operation. No notable differences were found between computer-assisted navigation and conventional implantation techniques as regards the rotational alignment of the femoral or tibial components. Although the deviation from the transepicondylar axis was relatively low, there was a considerable range of deviation for the tibial rotational alignment. There was no statistically significant difference regarding the occurrence pattern of outliers in mechanical malalignment but the number of outliers was reduced in the navigated group.
منابع مشابه
Alignment and orientation of the components in total knee replacement with and without navigation support: a prospective, randomised study.
Bilateral sequential total knee replacement was carried out under one anaesthetic in 100 patients. One knee was replaced using a CT-free computer-assisted navigation system and the other conventionally without navigation. The two methods were compared for accuracy of orientation and alignment of the components. There were 85 women and 15 men with a mean age of 67.6 years (54 to 83). Radiologica...
متن کاملEmbolic phenomena during computer-assisted and conventional total knee replacement.
Systemic emboli released during total knee replacement have been implicated as a cause of peri-operative morbidity and neurological dysfunction. We undertook a prospective, double-blind, randomised study to compare the cardiac embolic load sustained during computer-assisted and conventional, intramedullary-aligned, total knee replacement, as measured by transoesophageal echocardiography. There ...
متن کاملComparison of Clinical, Functional, and Radiological Outcomes of Total Knee Arthroplasty Using Conventional and Patient-Specific Instrumentation
Background: Recently, patient-specific instrumentation (PSI) systems have been developed in order to increase theaccuracy of component positioning during total knee arthroplasty (TKA); however, the findings of previous studies arecontroversial in this regard. In the current randomized clinical study, the outcomes of computer tomography (CT)-basedPS (patient specific)-guided TK...
متن کاملUse of the Cusum technique for evaluation of a CT-based navigation system for total knee replacement.
Most of the early failures of total knee replacements are related to technical flaws. Conventional ancillary devices achieve good alignment in the frontal plane in only an average of 75% of total knee replacements. Computer-assisted surgery may improve the technical quality of implantation surgery. The aim of our study was to evaluate the use of computer-assisted surgery using a quality control...
متن کاملImageless computer navigation in total knee arthroplasty.
OBJECTIVES To study the difference in clinical and radiological outcomes between imageless computer-navigated and the conventional technique for performing total knee arthroplasty. DESIGN Prospective case-control study. SETTING University teaching hospital, Hong Kong. PATIENTS Forty-seven patients with imageless computer-navigated total knee arthroplasty were matched with 47 patients with...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The Journal of bone and joint surgery. British volume
دوره 90 8 شماره
صفحات -
تاریخ انتشار 2008