Biomechanical analysis comparing three C1-C2 transarticular screw salvaging fixation techniques.
نویسندگان
چکیده
STUDY DESIGN This is an in vitro biomechanical study. OBJECTIVE To compare the biomechanical stability of the 3 C1-C2 transarticular screw salvaging fixation techniques. SUMMARY AND BACKGROUND DATA Stabilization of the atlantoaxial complex is a challenging procedure because of its complicated anatomy. Many posterior stabilization techniques of the atlantoaxial complex have been developed with C1-C2 transarticular screw fixation been the current gold standard. The drawback of using the transarticular screws is that it has a potential risk of vertebral artery injury due to a high riding transverse foramen of C2 vertebra, and screw malposition. In such cases, it is not recommended to proceed with inserting the contralateral transarticular screw and the surgeon should find an alternative to fix the contralateral side. Many studies are available comparing different atlantoaxial stabilization techniques, but none of them compared the techniques to fix the contralateral side while using the transarticular screw on one side. The current options are C1 lateral mass screw and short C2 pedicle screw or C1 lateral mass screw and C2 intralaminar screw, or C1-C2 sublaminar wire. METHODS Nine fresh human cervical spines with intact ligaments (C0-C4) were subjected to pure moments in the 6 loading directions. The resulting spatial orientations of the vertebrae were recorded using an Optotrak 3-dimensional Motion Measurement System. Measurements were made sequentially for the intact spine after creating type II odontoid fracture and after stabilization with unilateral transarticular screw placement across C1-C2 (TS) supplemented with 1 of the 3 transarticular salvaging techniques on the contralateral side; C1 lateral mass screw and C2 pedicle screw (TS+C1LMS+C2PS), C1 lateral mass and C2 intralaminar screw (TS+C1LMS+C2ILS), or sublaminar wire (TS + wire). RESULTS The data indicated that all the 3 stabilization techniques significantly decreased motion when compared to intact in all the loading cases (left/right lateral bending, left/right axial rotation, flexion) except extension. All the 3 instrumented specimens were equally stable in extension/flexion and lateral bending modes. TS+C1LMS+C2PS was equivalent to TS+C1LMS+C2ILS (P > 0.05) and superior to TS + wire in axial rotation (P < 0.05). Also, TS+C1LMS+C2ILS was superior to TS + wire in axial rotation (P < 0.05). CONCLUSION Fixation of atlantoaxial complex using unilateral transarticular screw supplemented with contralateral C1 lateral mass and C2 intralaminar screws is biomechanically equivalent to C1 lateral mass and C2 pedicle screws and both are biomechanically superior to C1-C2 sublaminar wire in axial rotation.
منابع مشابه
Posterior atlantoaxial fixation: biomechanical in vitro comparison of six different techniques.
STUDY DESIGN Six different techniques for atlantoaxial fixation were biomechanically compared in vitro by nondestructive testing. OBJECTIVE To evaluate the immediate three-dimensional stability of a new atlas claw combined with transarticular screws and alternative techniques for transarticular screw fixation in comparison with established techniques. SUMMARY OF BACKGROUND DATA Posterior tr...
متن کاملBiomechanical Analysis of Novel Occiput-C1-C2 Fixation Techniques
Introduction The etiology of occipitocervical instability is vast and includes trauma, rheumatoid arthritis, infection, tumor, congenital deformity, and degenerative processes. The development of instability in this region can result in intractable pain with or without neurological deficit and requires surgical treatment. Surgical management has evolved during the past decades, and currently, i...
متن کاملBiomechanical Comparison of Five Different Atlanto-axial Posterior Cervical Fixation
Introduction: Previous investigations have demonstrated that threepoint fixation, using bilateral transarticular screws in combination with posterior wiring, provide the most effective resistance to minimize motion about C1-C2. However, transarticular screws are technically demanding and require considerable experience. Posterior wiring techniques affording one point of fixation, such as Brooks...
متن کاملBiomechanical comparison of two stabilization techniques of the atlantoaxial joints: transarticular screw fixation versus screw and rod fixation.
OBJECTIVE To compare the biomechanical stability imparted to the C1 and C2 vertebrae by either transarticular screw fixation (TSF) or screw and rod fixation (SRF) techniques in a cadaver model. METHODS Ten fresh ligamentous human cervical spine specimens were harvested from cadavers. The specimens were tested sequentially in the intact state, after injury and stabilization (unilateral left si...
متن کاملAtlantoaxial fixation: overview of all techniques.
Over the past century, steady advances have been made in fixating an unstable atlantoaxial complex. Current options for fixation of the atlantoaxial complex include posterior clamps, posterior wiring techniques, C1-C2 transarticular screw fixation, posterior C1 lateral mass screw with C2 pars or pedicle screw fixation, and anterior transoral C1 lateral mass to C2 vertebral body fixation.
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Spine
دوره 35 4 شماره
صفحات -
تاریخ انتشار 2010